DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 24 OF 35
Figure 3-19: Screen Brightness Setup Figure 3-20: Battery Status info
3.5
TROUBLESHOOTING
Cause
Solution
Problem: The Infrascanner screen does not turn on when Disposable Fiber Optic
Shield is placed on.
There are no batteries installed.
Install either the rechargeable battery pack or use
fresh 4 AA disposable batteries.
The battery of the Infrascanner is
drained.
Recharge the Infrascanner or use fresh 4 AA
disposable batteries
The disposable batteries are
inserted incorrectly
Ensure that the batteries are installed with the
correct polarity as marked on the case.
Problem: The Infrascanner screen flashes or the Infrascanner continuously
restarts when the Disposable Fiber Optic Shield is placed on.
The battery of the Infrascanner is
low.
Recharge the Infrascanner or use 4 fresh AA
disposable batteries.
Problem: The Cradle amber Charge indicator does not illuminate when charging.
Disposable batteries are installed
in Infrascanner
The Cradle does not charge disposable AA
batteries.