18
Configuring
the
Analyzer
After plugging in the 8400D the Power Indicator will be illuminated confirming that there is power to the
analyzer - the power supply is properly attached or the unit’s battery is charged. Press the On/Off button
to activate the analyzer. During the initial power-up, the pump will come on briefly and there will be a
status bar at the bottom of the screen to indicate the calibration of O2. When this is complete, the Standby
Screen will appear. The operator MUST zero the analyzer using the corresponding soft key before
exhaust gas testing. Refer to Chapter 3 “Display Features” if you are not sure which button to use.
S T A N D B Y
The 8400D will return to this screen after exiting the
menu selection. It is possible to customize the display
screen settings by going to Display Settings in Menu 2.
This allows the operator to select one or as many as all
of the values to be displayed on the screen.
M E A S U R E
– Allows manual on/off operation of the pump. This function can also be used to purge gas
from the analyzer. For more detailed information, please refer to the section “Identifying Actions When
Pump Is On”.
Z E R O
– Zeros the analyzer for approximately 1 minute. The pump is automatically controlled during
this phase.
P L A Y
– Plays back recorded data. Choose Pause/Continue to control the play function or Exit to return
to the Standby screen.
M E N U
– Use this button to access Menu 1
Содержание IR-8400D
Страница 1: ...IR 8400D Continuous Gas Monitoring Analyzer ...
Страница 2: ......
Страница 4: ...3 ...
Страница 6: ...5 ...
Страница 8: ...7 ...
Страница 12: ...11 ...
Страница 14: ...13 ...
Страница 16: ...15 Figure 2 IR 8400D Outlying Dimensions ...
Страница 28: ...27 ...
Страница 36: ...35 ...