I D E N T I F Y I N G A C T I O N S W H E N P U M P I S O N
The 8400D will display this screen when
Measure
is
selected from the Standby screen. It is possible to
customize the display screen settings by going to
Display Settings in Menu 2. This allows the operator to
select any or all of the values shown to be displayed on
the screen.
M E A S U R E –
Selecting
MEASURE
allows the operator to turn the pump on and off. When the pump
is off, the display will return to the Standby screen. To start pump when the measurement of a specific
gas concentration is desired simply press the
MEASURE
button. Once the pump has been turned on the
screen menu will change from
MEASURE
,
ZERO
,
PLAY
, and
MENU
, to
MEASURE
,
,
HOLD
, and
RECORD
. Pressing the
MEASURE
key once more while the pump is running will stop the pump and
return you to the standby menu.
P R I N T –
Prints data to a printer connected to the serial port located in the Junction Box.
H O L D / R E S U M E –
Selecting the
HOLD
key will lock the display values to capture that moment in
time. Once the
HOLD
key has been initiated the
HOLD
enunciator will change from
HOLD
to
RESUME
.
To return to normal operation press the
RESUME
key and return to a live display.
R E C O R D –
The analyzer has the ability to store data by selecting
RECORD
. The recorded
information will remain in memory until it’s erased. To erase the data in memory enter into the Menu 1
options and select Erase Data. Once you have selected this option you will be asked if you wish to erase
stored data. Select
YES
to erase data or
NO
and return to the standby screen.
Содержание IR-8400D
Страница 1: ...IR 8400D Continuous Gas Monitoring Analyzer ...
Страница 2: ......
Страница 4: ...3 ...
Страница 6: ...5 ...
Страница 8: ...7 ...
Страница 12: ...11 ...
Страница 14: ...13 ...
Страница 16: ...15 Figure 2 IR 8400D Outlying Dimensions ...
Страница 28: ...27 ...
Страница 36: ...35 ...