Figure 37 Make sure transmitter seated in the sensor base.
Figure 38 Gently peel the tape off the skin.
Removed the sensor together with transmitter. Do not
remove the transmitter when the sensor base is attached to the
skin.
After the sensor base is peeled off, the transmitter can be
removed. Remove the transmitter using any of the following
two methods:
Method 1: the safety lock can be used as a tool to remove
the transmitter.
1. Place the sensor base/transmitter on the plane.
2.Hold the round edge of the safety lock.
3. Ensure that the serrated edge of the safety lock is
downwards, as shown in the figure below:
11.3 Remove the transmitter
12 Faults and Troubleshooting
This chapter provides effective advice and instructions
for using the continuous blood glucose monitoring system.
12.1 Sensor insertion failure
Figure 39 Removed transmitter with safety lock
Method 2 : open the buckles on both sides of the
sensor base and remove the transmitter by fingers.
12.1.1 Failure to pull out the safety lock
Make sure the safety lock is pulled away from the body.
Use the arrow as a guide to remove the safety lock.
12.1.2 Failure to pull out the guide seat failure
Make sure the plunger is fully pushed to the bottom before
removing the guide seat.
Pull out the guide seat vigorously.