15
Wash your hands thoroughly and dry them.
Remove the sensor from the package and check it
carefully to make sure it is intact.
The applicator is a disposable assistive device used alone.
Safety lock can prevent accidental release of the sensor
electrode before you are ready.
5.3 Selection of insert site
Caution
It is forbidden to insert sensors in areas that may be
crush-squeezed, tattooed, or allergic. These sites are not
ideal for measuring blood glucose.
It is forbidden to insert sensors within 8 cm of insulin
injection and placement of insulin pump.
Performance of the sensor may be affected when
inserted in those areas, resulted in missing low/high
blood glucose levels.
It is recommended to select a position other than 5 cm
from the navel in the abdomen where the sensor is to be placed.
The position of the sensor is generally selected to be flat, not
easily deformed, and not easily rubbed by a belt or a seat belt.
It is forbidden to insert sensors within 8 cm of insulin
injection and placement of insulin pump.
The area where put the sensor needs to be shaved and
cleaned so that the adhesive tape is kept firm.
Change the insertion site each time.
Make sure there are no traces of lotions, perfumes, or
drugs on the sensor.
5.4 Place sensor
Caution
Continuous blood glucose testing cannot be used as
a basis for diabetes diagnosis, such as insulin injection.
It cannot replace the blood glucose meter. The value of
blood glucose may not be consistent with the reading of
blood glucose. Using glycemic readings as a diagnostic
basis for diabetes may result in low/high blood glucose.
1.Before inserting the sensor, clean the skin with a traditional
method (such as an alcohol tablet) and let it dry.
2.Remove the protective paper from the bottom of the s
ensor base. Hold the sensor by the applicator barrel, and don
not touch the adhesive tape.