14
5 Sensor insertion and startup
Blood glucose meters and test strips are not included in
the continuous blood glucose test system. When calibrating,
you need to prepare a blood glucose meter and a test strip.
The sensor continuously monitor displays blood glucose
readings for 14 days. The following steps will show you how
to insert the sensor.
Please read the instructions.
Warning
In rare cases, the sensor electrode may fracture.
Don't ignore it. If the sensor electrode breaks and there
is no visible sensor electrode on the skin, do not
attempt to remove it. Seek professional medical help in
the condition of infection symptoms or inflammation -
redness, swelling or pain. If the sensor electrode
ruptures, please report to our technical support
department.
No safety hazard was found in the patient's
residual rupture sensor electrode during in vitro NMR
test. There was no obvious migration or heating of the
sensor electrode, and the imaging was limited to the
area around the sensor electrode.
5.1 Preparation
Check the expiry date on the sensor package. insert the
sensor before the expiration date.
Warning
The storage temperature of the sensor is 2oC-
25oC, which can be transported at room temperature.
The storage time is the lifetime (8 months) of the
sensor. If the temperature range of the refrigerator is
2oC-25oC, the sensor can be stored in the refrigerator.
Sensors should not be stored in the freezer. Improper
storage results in inaccurate blood glucose readings
and missing low/high blood glucose levels.
Caution
If the sensor is stored in the refrigerator, please
take it out half an hour before use. The sensor can
be inserted at room temperature.
5.2 Remove the sensor
Caution
Do not use sensors with damaged or opened
sterile packaging. Using an unsterilized sensor can
cause infection.