7 | Warning and error messages
INFICON
82 / 108
XL3000flex-Operating-instructions-jina83en1-07-(2105)
Warning
(Wrn)
Error (Err)
Error message
LDS3000
Error number
Limit values Cause
LDS1000
Protocol
Binary or
ASCII
protocol
compatibilit
y mode
LDS1000/
LDS2010
Err306
Anode voltage faulty
36
127
40 V
deviation
from the
default
value
The anode voltage does not
match the default value or the set
value is outside the allowable
setting range.
Wrn310
Cathode 1 is defective
45
136
Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Wrn311
Cathode 2 is defective
46
137
Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Err312
Cathode defective
47
138
Cathode defective, line to cathode
interrupted, IF board or MSB
defective
Err340
Emission error
44
135
< 90% of
the target
value
> 110% of
the target
value
Emission was stable previously,
pressure probably too high,
message after 15s
Wrn342
Cathode not connected
47
138
Both cathodes defective during
self-testing after switch on or plug
not connected
Wrn350
Suppressor not connected
39
130
Supressor cable is not plugged in
or defective during the self-test
after switching on
Wrn352
Preamplifier not connected
Preamplifier defective, cable not
plugged in
Err358
Preamplifier oscillates
between 2 ranges
Signal varies too much (see
command 1120)
Preamplifier defective
Wrn359
Overdriven preamplifier
31
123
Signal too large preamplifier
defective
Содержание XL3000flex
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 92: ...9 Decommissioning the measuring instrument INFICON 92 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Страница 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 107: ......
Страница 108: ......