Warning
(Wrn)
Error (Err)
Error message
LDS3000
Error number
Limit values Cause
LDS1000
Protocol
Binary or
ASCII
protocol
compatibilit
y mode
LDS1000/
LDS2010
Wrn360
Preamplifier output too low 31
123
<-70 mV at
500 GΩ
Poor ion source or contaminated
mass spectrometer
Wrn361
Preamplifier offset too high 31
123
>+/-50 mV
at 500 GΩ,
>+/-10 mV
at 15 GΩ,
<+/-10 mV
at 470 MΩ,
<+/-9 mV
at 13 MΩ
Preamplifier defective
Wrn362
Preamplifier range error
31
123
Preamplifier or MSB box defective
Wrn390
500 G outside the range
31
123
450 GΩ
550 GΩ
Preamplifier defective, error at the
suppressor, IF board or MSB
defective
4xx TMP fault (also temperature)
Err400
TMP fault number
49
15
Wrn401
TMP warning number
Err402
No communication with
TMP
49
15
Cable to TMP / TMP defective, IF
board or MSB defective
Err403
TMP rotation speed too
low
53
142
< 95% of
the target
value
Pressure too high, TMP defective
Err404
TMP current consumption
too high
49
2
3A
Err405
No TMP run-up time
60
61
5 min.
Pressure too high, TMP faulty
Err410
TMP temperature too high
49
2
61°C
Cooling failed, check MSB module
operating conditions
Wrn411
High TMP temperature
49
2
60°C
Cooling failed, check MSB module
operating conditions
Err420
TMP voltage too high
49
2
Power supply defective, TMP
defective
Содержание XL3000flex
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 92: ...9 Decommissioning the measuring instrument INFICON 92 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Страница 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 107: ......
Страница 108: ......