6 | Operation
INFICON
70 / 108
XL3000flex-Operating-instructions-jina83en1-07-(2105)
– Generally select signal filter I-Filter for the operation mode "Sniff".
– If the signal filter should simulate the time behavior of older units, then select filter
"Fixed" or "2-Zone".
I•CAL
The leak rates are averaged at time intervals that are optimized
for the range of the leak rates. The algorithm used offers
excellent sensitivity and response time. Use of this setting is
strongly recommended.
fixed
The leak rates are averaged at fixed intervals of 0.2 seconds.
2-zone
The filter is compatible with LDS1000 and LDS2000. The
averaging period is switched depending on the filter leak rate
threshold.
I-Filter
Filter optimized for sniffer mode.
(Default with XL Sniffer Adapter set)
I-Filter slope
suppress.
Same as I-Filter, but with additional slope suppression. The
edge suppression corrects the measurement changes during
the warm-up phase.
Control unit
Main Menu > Settings > ZERO/Filter > Filter > Filter mode
LD protocol
Command 402
ASCII protocol
Command *CONFig:FILTER
Setting the filter leak
rate threshold
Leak rate background in mbar l / s for the averaging period. The averaging period is
10.24 s below this value. Above this value, the averaging period is 160 ms. Setting
applies only to filter "2-stage".
1E-11 ... 9.9E-3
Control unit
Main Menu > Settings > ZERO/Filter > Settings > Filter 2-zone
LD protocol
Command 403
ASCII protocol
Command *CONFig:LRFilter
Setting filter ZERO
time
Update interval for the offset value with negative leak rate signal (except for I•CAL
filter).
Resolution 0.1 s (50 = 5.0 s)
Control unit
Main Menu > Settings > ZERO/Filter > Settings filter > ZERO
time
LD protocol
Command 411
ASCII protocol
Command *CONFig:ZEROTIME
Содержание XL3000flex
Страница 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Страница 48: ...6 Operation INFICON 48 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 92: ...9 Decommissioning the measuring instrument INFICON 92 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 103: ...INFICON RoHS 13 XL3000flex Operating instructions jina83en1 07 2105 103 108 13 RoHS ...
Страница 106: ...Index INFICON 106 108 XL3000flex Operating instructions jina83en1 07 2105 ...
Страница 107: ......
Страница 108: ......