Trigger & Alarms
•
>
The trigger levels, the volume of the loudspeaker and the units of leak rates and
pressures can be set in this menu
6.4.1
Trigger Level 1
•
>
>
The value of the first trigger level can be set. See Chapter
for the description of the entry.
Trigger 1 and Trigger 2 are programmable switching thresholds. When these
thresholds will be exceeded the UL1000 and UL1000 Fab reacts as follows:
Display
In the status line of the display the signs for Trigger 1 and Trigger 2 are displayed
inverted if the leak rate exceeds (becomes higher than) the programmed value.
Relay Output
The trigger-relais of the digital out switches. Please refer to Chapter
, Digital
out, for further details.
Alarm/Loudspeaker
Additionally Trigger level 1 defines at which level the various alarm types react (See
Chapter
,
)
Fig. 6-9: The
Menu
Key No.
Name
Description
1
Back
Return to the main menu.
2
Definition of Trigger level 1. See chapter
3
Definition of Trigger level 2. See chapter
4
See chapter
5
Selection of leak rate and pressure units.
See chapter
6
Not used in this menu.
7
See chapter
8
Choice of different alarm types. See chapter