Equipment Settings
73
4.5.5
Baud rate & end sign
The baud rate can be set to values between 1200 and 19200. Default setting: 9600.
The end sign can be set to Carriage Return (CR), Line Feed (LF) or to CR+LF. This
setting is only relevant for data output of the P3000. For input strings always CR has
to be used.
Default setting: CR+LF.
4.5.6
PRO-Check
In this submenu the PRO-Check reference leak can be disabled completely so that
for Protec P3000 leak detectors without a PRO-Check onboard the leak detector
does not check for the electrical connection with the reference leak anymore and no
error message is issued.
Default value: Enable
4.6
The Info Menu
The
INFO
MENU
lists all internal data that may be useful for any trouble-shooting of the
Protec P3000. The
INFO
MENU
consists of 10 pages. On each page the upper left
button “
BACK
” allows to leave the info menu, the lower left button allows to go back
one page whereas the lower right button allows to go to the next page. The page
number is displayed in the upper right corner.
Fig. 58
Customizing PLC inputs
Menu item
Format
Description
Page 1: General Data
Foreline pressure
mbar
Flow
sccm
Flow through sniffer line
Time since power on
Min