Equipment Settings
When set to IꞏGuide mode the PRINTER MANUAL mode is without function.
Printer auto
In
PRINTER
AUTO
the leak rate will be sent any time the set trigger level is exceeded.
At the time the signal drops below the trigger level the maximum detected leak rate
will be sent.
When set to IꞏGuide mode the leak rate results will be sent after each test point
together with the test point number. After the test cycle has been completed, the
summarized global leak rate will be sent.
4.5.4
Select PLC inputs
In the
SELECT
PLC
INPUTS
submenu the user can select which pin on the I/O port
(suitable for PLC input) represents which command. The default setting is as follows:
To change these settings select the appropriate pin with the UP and DOWN arrows
on the left side of the display and afterwards select the desired command from the
list of commands with the UP and DOWN buttons on the right side of the display.
Press OK to save your settings.
Format
Date
Time
Gas equivalent Leak rate
Unit
Example
03.05.2005
12:31
He
6x10
-4
mbar l/s
Format
Date
Time
Gas equivalent Leak rate
Unit
Example
03.05.2005
12:31
R134a
2.34
g/a
Example:
Point no.
Date
Time
Leak rate
Unit
Point 01
8.3 E-6
mbar l/s
Point 02
2.0 E-6
mbar l/s
Point 03
8.4 E-6
mbar l/s
Point 04
1.2 E-6
mbar l/s
Global
13.04.2007
11:57:03
1.8 E-6
mbar l/s
Pin
Default Command
Pin
Default Command
7
Standby
13
Cal
8
Clear Error
20
Zero
9
Cal Abort
25
Not used
Содержание Protec P3000
Страница 2: ......
Страница 10: ...10 General Safety Precautions...
Страница 121: ...CE Declaration of Conformity 121 9 CE Declaration of Conformity Fig 91...
Страница 122: ...122 CE Declaration of Conformity...
Страница 125: ...Appendix 125...
Страница 126: ......
Страница 127: ......
Страница 128: ......