Equipment Settings
65
Peak hold
The
PEAK
HOLD
function can be enabled or disabled in this submenu. If the
PEAK
HOLD
function is enable the maximum leak rate will be displayed for a preset time in addi-
tion to the currently measured value in the main display. The time setting will have
no effect if this function is disabled.
Default: value disabled, 5 s.
Flow display
In the
FLOW
DISPLAY
submenu the continuous display of the flow through the sniffer
line during measurements can be
ENABLED
or
DISABLED
. If
ENABLED
, the flow will be
indicated in the
STANDARD
OPERATION
MODE
only underneath the leak rate bar graph
as integer number.
Notice
After certain events like acknowledging warnings or when switching the
flow mode, the flow monitoring may be disabled for a short period of time
while the flow stabilizes to a final value again and no new warnings will be
issued. During this time a message ”Flow control inactive” will be displayed
instead of the flow value.
Default setting: disabled
Fig. 54
Setting-up the contrast of the main display
Fig. 55
Setting-up the peak hold function