
Operation of the Protec P3000
43
By pressing the right probe button the next testing cycle may be started.
Notice
The I•Guide mode can also be used as a timer signal only. If the number of
points is set to 0, the Protec P3000 will prompt for the next location to be
tested continuously without using the global leak rate function.
Notice
The I•Guide mode can also be used to summarize leak rates on demand.
If the number of points is set to 99, a results screen with summarized global
leak rate will be displayed after the right button has been pressed for 2 s
conintously (or after the 98th point automatically).
3.4.3
The Info Page
By pressing the info button on the main display a general info page will open. Infor-
mation on the software version used, date & time information, the currently set audio
volume and the minimum audio volume are stated.
If an active warning exists this will be displayed in the info page instead of the gas
info lines.
Fig. 29
Result of I•Guide program: unit under test failed
Fig. 30
Info page without errors or warnings
Содержание Protec P3000
Страница 2: ......
Страница 10: ...10 General Safety Precautions...
Страница 121: ...CE Declaration of Conformity 121 9 CE Declaration of Conformity Fig 91...
Страница 122: ...122 CE Declaration of Conformity...
Страница 125: ...Appendix 125...
Страница 126: ......
Страница 127: ......
Страница 128: ......