24
tira85e1-a (2014-10) PxG55x.cp
Index
SI
DataType
NV Access PM Name
8001
0x01 REAL
x
RW
Zero Adjust Mode
The CDG Sensor is adjusted automatically by the Pirani sensor, if this index is set.
The High Trip Point is calculated by:
High Trip Point = Value
*)
× Percentage
**)
*)
Value from Piezo Sensor (I: 0x6010, SI: 0x11)
**)
Percentage
High Trip Source (I: 0x800E/F, SI: 0x13)
In this case, the High Trip Point value is automatically updated by the device itself.
Any user-written value to the parameter High Trip will be refused with an abort
code.
The Low Trip Point is a fixed value:
The value defined in Low Trip Point Limit is compared with the pressure value
referenced by the Source Index parameter.
High Trip Point
Time t
in-
active
active
in-
active
active
Low Trip Point
Measurement signal
(Pressure p)
Low Trip
Hysteresis
High Trip
Hysteresis
in-
active
Index
SI
DataType
NV Access PM Name
800E
0x01 BOOL
x
RW
High Trip Enable
0x02 BOOL
x
RW
Low Trip Enable
0x11 REAL
x
RW
High Trip Point Limit
0x12 UDINT
x
RW
High Trip Source Index
0x13 REAL
x
RW
Percentage High Trip
Source
0x14 REAL
x
RW
Low Trip Point Limit
0x15 UDINT
x
RW
Low Trip Source Index
0x16 REAL
x
RW
Percentage Low Trip
Source
0x17 REAL
x
RW
High Trip Hysteresis
0x18 REAL
x
RW
Low Trip Hysteresis
0x1A
UDINT
x
RW
Source
Index
4.7.2 Configuration
Capacitance Diaphragm
Subindex 0x01
4.7.3 Configuration Trip
Point 1