tira85e1-a (2014-10) PxG55x.cp
17
Safe Value: The value to be used for Safe State = Safe Value.
•
PCG gauges with ATM: The "Manufacturer Configuration Module 3" is assigned
to the Pirani module.
Index
SI
DataType
NV Access PM Name
4020 0x01
BYTE RW Safe
State
0x02
REAL
RW
Safe
Value
Specifies the behavior for the Value for states other than Valid.
Safe State
0
1
2
3
Zero
Full Scale
Hold last value
Use safe value
Safe Value: The value to be used for Safe State = Safe Value.
Index
SI
DataType
NV Access PM Name
6000 0x0E
BOOL RO tx
TxPdoState
0x11
REAL
RO
tx
Sensor
Value
Is set if the device is not in Safe State (value (I 0x6nn0, SI 0x11) = valid)
TxPdoState
0
1
Invalid
Valid
The corrected, converted, calibrated final analog input value of the sensor.
Index
SI
DataType
NV Access PM Name
6001 0x01
BOOL RO tx
Reading
Valid
0x02
BOOL
RO
tx
Overrange
Exceeded
0x03
BOOL
RO
tx
Underrange
Exceeded
Indicates whether the Value parameter contains a valid value within the specified
accuracy or not.
Reading Valid
0 Invalid
1 Valid
Subindex 0x02
4.3.3 Manufacturer
Configuration Module 3
(PCG with ATM only)
Subindex 0x01
Subindex 0x02
4.4 Input Area PCG with
ATM (0x6000…0x6FFF)
4.4.1 Input Common
Subindex 0x0E
Subindex 0x11
4.4.2 Input Capacitance
Diaphragm
Subindex 0x01
Содержание PCG550
Страница 59: ...tira85e1 a 2014 10 PxG55x cp 59 Notes ...