Data Logging
The MX4 performs continuous data logging at 10 second intervals and
can hold approximately three month’s of data for the four-sensor
configuration. Readings are time-stamped and are stored in a first-in-first-
out (FIFO) queue that is overwritten when full. In addition to data logging,
the MX4 has an event log that records the last 15 alarm events.
Data can be downloaded to a PC via the software package and
Charger/Datalink. Data is extracted from the MX4 via a Charger/Datalink.
To purchase a Charger/Datalink, please contact your local distributor of
Industrial Scientific products.
To use the Charger/Datalink, you must first install the setup software
located on the CD that is supplied with the Charger/Datalink. Connect the
Charger/Datalink to a USB port on your PC, click on the “Connect” button
to establish communications. Once communications are established, data
can be downloaded or cleared from the interface menu. To view data,
select “File Open”. To view graphics, select “Graphics” from the
spreadsheet menu. To disconnect at anytime, click on the “Disconnect”
button and remove the MX4 from the Charger/Datalink.
NOTE:
When a sensor is in an over-range condition, the maximum range
value will be logged into the datalog memory for that sensor.