Enter Zero Screen
The Enter Zero screen displays the zero
icon. This screen is used to launch the
zeroing process (explained earlier in this
manual, refer to calibration).
To begin the zeroing process, press the
Enter button. Press the On/Off/Mode button
to step to the calibration configuration
screen.
Calibration Mode Screen
The Calibration Mode screen displays the
gas bottle icon, the up and down arrow
icons, and the “CAL” text with either a “0”
(Standard-Cal mode) or a “1” (Quick-Cal
mode).
Use the Enter button to toggle between
values 0 and 1. After the desired calibration
mode (0 or 1) is selected, press the
On/Off/Mode button to display the Low
Alarm Setpoint screen
Low Alarm Set Point
The Low Alarm Set Point screen displays
the four low alarm set points, the alarm
icon, and the down arrow (low limit) icon.
If no changes are needed, press the
On/Off/Mode button to move to the next
screen (the High Alarm Set Point screen).
If changes are desired, press the Enter button. The first low alarm value
flashes. To adjust this value, use the Enter button. Once the desired value
is met, press the On/Off/Mode button to submit the value and select the
next low alarm value. Continue this process until all four low alarm set
points have been set.
Once all four values are set, press the On/Off/Mode button to submit the
values. Pressing the On/Off/Mode button again displays the High Alarm
Set Point screen.