3. Standing Posture
(1) It is recommended that the examinee stand for about 10~15 minutes before the test so that body water
may be dispersed evenly inside the body.
(2) Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity.
(3) Make sure arms do not touch the torso. Spread them naturally to a 15 degree angle away from trunk.
(4) Arms should be extended naturally.
(5) Make sure thighs do not touch each other, and spread legs shoulder-width apart.
Attention
• It is recommended that the examinee stand for about 10~15 minutes.
• Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity.
• For an examinee with high foot arches feet, the foot electrodes might not attach properly when using the
Touch Type electrodes. Make sure to attach them as best as possible.
• If examinees have dry hands and feet, wipe the equipment with an InBody tissue before testing.
• Due to the weight of the cable, it is easy for the Adhesive Type electrodes to fall off during the test.
Please be careful and make sure that the electrodes are attached properly.