D. Test Posture
In order to use the BWA2.0 correctly, it is necessary to understand the correct measuring posture.
Maintain consistent measurement posture for high reproducibility and reliable results.
1) Lying Posture
Accurate posture
Measurement Posture
•
The BWA2.0 is located next to the knee as shown above.
•
Open your arms slightly beyond 15 degrees to keep your arms and torso from touching.
•
Keep your legs slightly open to shoulder width to prevent thighs from touching.
• Measure after lying down for at least 10-15 minutes.
• Be careful not to touch the electrode cable and body parts with the frame when measuring on a bed with an iron frame.
• If there is a heating mat (electric blanket) on the floor or mat, be sure to turn it off and unplug the power cable if possible.
• Disposable electrode (BWA-ES100) may detach due to the weight of the electrode cable when you are using forceps
electrode. Please be careful.
Caution
•
Allongez-vous durant au moins 10-15 minutes avant de prendre les mesures.
• Évitez tout contact entre un cadre de lit métallique (s'il y a lieu) et le câble de l'électrode et/ou les différentes parties du corps.
• Si le sol ou le tapis est recouvert d'un tapis chauffant (couverture électrique), assurez-vous de l'éteindre et, si possible, le
débrancher.
• L'électrode jetable (BWA-ES100) tombe facilement à cause du poids du câble de l'électrode lorsque vous utilisez l'électrode
forceps.
Veuillez faire attention.
Attention