If a problem persists, please contact Product Support.
A. Error Code
The error messages will display on the screen if they occur during the test.
Please provide the error code to Product Support.
B. Regarding the Equipment
If your technical issues cannot be resolved with the information below, please contact Product Support.
Place your elbow on the elbow point.
Press the START button.
Do not move during the measurement.
Irregular signal was detected.
SYS.
P.R
DIA.
IV.
Issue
Potential Solution
· The device is not
turning on.
·
Plug the power cable into a properly
grounded (3-prong) outlet.
·
When using a power strip,
make sure the switch is turned on.
·
Make sure you are using the power cable (AC 250V 10A)
provided by InBody.
·
Issue may occur when the fuse is blown. Replace the fuse by following the
instructions below:
The fuse socket is located on the back of the
device beneath the power plug.
Pull the fuse out by using a flathead screwdriver. Replace
the fuse with the provided spare fuse. Additonal fuses can
be purchased if necessary.
3-prong outlet
Power switch of
power strip
1
2
<110V>
<220V>