D. Product Specifications
Test Instructions
Oscillometric
Test range
Pressure: 0 to 300 mmHg
Pulse: 30 to 240 bpm
Degree of precision
Pressure: ± 2mmHg
Pulse: Within ± 1.5%
Test result
Systolic blood pressure, diastolic blood pressure, pulse rate, pulse pressure,
mean arterial pressure, rate pressure product
Test Time
Approx. 30 seconds on average
(20-50 seconds depending on the pulse and blood pressure value)
Pressurization Time
Approx. 10 seconds
Minimum scale unit
1mmHg
Screen Type
7-Segment LED
Results Sheet outputs
Systolic and Diastolic blood pressure, Mean arterial pressure, Pulse pressure, pulse,
Rate pressure product, Pulse envelope, Blood pressure graph, Check posture, and
Blood pressure knowledge
Storage function
Can store the total results for up to 1 million tests
Energy saving
Automatically enters the energy saving mode after 2 minutes of the idle time
Dual safety
mechanism
The air will be released when the Start/Stop button is pressed.
The air will be released automatically when the air pressure exceeds 300 mmHg.
The air will be released when the Emergency Stop button is pressed.
(The dual safety mechanism enables the cuff to release the air regardless of the
central controller when the Emergency Stop button is pressed.)
Voice guidance
Guides user through blood pressure measurement and results
Printer
High-speed thermal printer with a built-in automatic cutter (width: 2.5 inch/6.35 cm)
Rated voltage and
Power consumption
AC 100-120 / 200-240V, 50/60Hz, 36VA
Dimensions
Approx. 299 (W) × 547 (D) × 485 (H) mm
Approx. 11.8 (W) × 21.5 (D) × 19.1 (H) in
Weight
Approx. 7.1 kg (15.65 lb)
Operating environment
Temperature 10 ~ 40ºC (50 ~ 104ºF) , Humidity 30 ~ 75% RH, Atmospheric pressure
70 ~ 106 kPa
Storage environment
Temperature -10 ~ 70ºC (14 ~ 158ºF), Humidity 10 ~ 80% RH,
Atmospheric pressure 50 ~ 106 kPa (No Condensation)
*
Specifications are subject to change without ntoice for purposes of product improvement.