Troubleshooting
In addition to the following troubleshooting items, the In-Situ website includes instructional videos, technical notes, and
more. See
www.in-situ.com/
.
Problem
Possible Cause
Possible Solution
Win-Situ 5 Software cannot
connect to the instrument.
Wrong COM port is selected,
communication settings are incompatible,
cable connections are loose or dirty, or
batteries are low.
Make sure cable connections are tight and
connectors are clean and dry.
Make sure the cable is securely attached to the
instrument.
Make sure the correct COM port is selected (select
Preferences
>
Comm Settings
to verify).
Make sure the communication settings in Win-Situ
and in the instrument match. To reset the device
communication settings to the serial defaults, click
“Reset all Devices” in the Comm Settings dialog
(
Preferences
>
Comm Settings
).
Make sure the internal battery has voltage
remaining, or external power is supplied.
See
www.in-situ.com/Technical_notes
and scroll
to the Communications and Software Technical
Notes section for more information.
Real-time readings display the
wrong units.
Default units are being used.
Click the
Sensors
tab, select the sensor, click the
configure button and select the desired units for
each parameter in the Sensor Setup window.
Cannot add a new log.
Only one active log can reside in the
device at a time—an active log is a log
that is Ready, Pending, Running, or
Suspended as shown in the Status
column of the Logging Tab.
The device has its maximum number of
logs already stored—the Level TROLL
300, 500, and Baro TROLL have a
capacity of 2 logs.
Stop or delete the log if possible. Alternatively,
configure the new log after the active log is
completed.
Download, and then delete a log you no longer
need. This will make room for an additional log on
the device.
New log exceeds available
memory (software message).
The log as configured would exceed the
device memory.
Edit the log and select a longer sampling interval.
If available, select the "wrap data" option. This
causes more recent data to overwrite older data
when memory is full.
For a log with a scheduled start, select None as the
stop condition, or select a stop time that is closer to
the start time.
Содержание Level TROLL 400
Страница 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Страница 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Страница 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Страница 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Страница 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Страница 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Страница 81: ...1 800 446 7488 81 www in situ com Declarations of Conformity and Similarity Level TROLL 400 Instrument ...
Страница 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Страница 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Страница 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Страница 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...