Level TROLL 700 Instrument
General
Level TROLL 700
Temperature ranges
1
Operational: -20-80° C (-4-176° F)
Storage: -40-80° C (-40-176° F)
Calibrated: -5-50° C (23-122° F)
Diameter
1.83 cm (0.72 in.)
Length
21.6 cm (8.5 in.)
Weight
124 g (0.27 lb)
Materials
Titanium body; Delrin nose cone
Output options
Modbus/RS485, SDI-12, 4-20 mA
Battery type & life
2
3.6 V lithium; 10 years or 2M readings
External power
8-36 VDC
Memory
Data records
3
Data logs
4.0 MB
260,000
50
Log types
Linear, Fast Linear, Linear Average, Event, Step Linear, True Logarithmic
Fastest logging rate & Modbus rate
4 per second
Fastest SDI-12 & 4-20 mA output rate 1 per second
Real-time clock
Accurate to 1 second/24-hr period
Sensor Type/Material
Piezoresistive; titanium
Calibrated Range
(Usable Depth)
Absolute (non-vented)
30 psia (11 m, 35 ft)
100 psia (60 m, 197 ft)
300 psia (200 m, 658 ft)
500 psia (341 m, 1120 ft)
1000 psia (693 m, 2273 ft)
Gauged (vented)
5 psig (3.5 m, 11.5 ft)
15 psig (11 m, 35 ft)
30 psig (21 m, 69 ft)
100 psig (70 m, 231 ft)
300 psig (210 m, 692 ft)
500 psig (351 m, 1153 ft)
Burst pressure
Max. 2x range; burst > 3x range
Accuracy (FS)
4
±0.05%
Resolution
±0.005% FS or better
Long-term stability
5
< 0.1% FS
Units of measure
Pressure: psi, kPa, bar, mbar, mmHg, inHg, cmH
2
O, inH
2
O
Level: in, ft, mm, cm, m
Temperature Sensor
Accuracy & resolution
±0.1° C; 0.01° C or better
Units of measure
Celsius or Fahrenheit
Warranty
3 years
Extended warranties are available for all instruments—call for details
Footnotes
See See page 8.
Содержание Level TROLL 400
Страница 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Страница 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Страница 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Страница 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Страница 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Страница 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Страница 81: ...1 800 446 7488 81 www in situ com Declarations of Conformity and Similarity Level TROLL 400 Instrument ...
Страница 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Страница 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Страница 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Страница 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...