Cleaning and Storage
Cleaning the Instrument
Clean the instrument body with water and a soft brush or plastic scouring pad, or soak overnight in a mild acidic solution,
such as household vinegar.
NEVER
submerge the connector portion of the instrument when it is not connected to a cable.
If the ports near the pressure sensor are clogged with silt or mud, try the following procedures.
l
Agitate the instrument vigorously in a bucket of clean water.
l
Apply a gentle rinse of water from a wash bottle.
l
In severe cases, remove the nose cone and clean out the holes with a soft brush or pipe cleaner.
Do not attempt to remove material from the instrument by tapping the instrument against a surface. To avoid damage to the
pressure sensor diaphragm, do not insert any object into the sensor opening or attempt to dig out dirt or other materials.
You
void the instrument's warranty by inserting anything into the sensor opening.
If contamination cannot be removed
using the recommendations above, please contact In-Situ for cleaning.
A TROLL pressure sensor with the nose cone removed. Do not insert
ANY
object into the sensor opening. Doing so voids the
warranty.
Twist-Lock Connectors
Keep the pins on all connectors free of dirt and moisture by using the soft protective dust caps when cable is not attached.
Storage
Store the instrument in a clean, dry place. Place the protective red dust cap on the cable end or store with cable attached to
protect the connector pins and O-ring. For vented cables, ensure the desiccant used is the appropriate size and change it
when needed. Store the instrument where it will not roll off a bench onto a hard surface or sustain other mechanical shock.
Protect the instrument from temperature extremes using the following guidelines:
l
Level TROLL Instruments—store within the temperature range -40° C to +80° C (-40° F to +176° F)
Содержание Level TROLL 400
Страница 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Страница 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Страница 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Страница 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Страница 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Страница 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Страница 81: ...1 800 446 7488 81 www in situ com Declarations of Conformity and Similarity Level TROLL 400 Instrument ...
Страница 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Страница 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Страница 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Страница 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...