Power Connections
The Red wire provides power for Modbus and SDI-12 modes. The Brown wire provides power for the 4-20 mA mode. If power
is present on the Brown wire and not on the Red wire, the device enters the 4-20 mA mode automatically and stays in the 4-
20 mode until power is removed from the Brown wire or is applied to the Red wire. The Red wire has priority. If power is
applied to the Red and Brown wires at the same time, the device will operate in Modbus or SDI-12 modes but not in 4-20.
Communication Protocols
The device automatically switches between Modbus and SDI-12 modes depending on which of the two interfaces has
activity. Modbus and SDI-12 cannot be used at the same time. The communication protocol that is currently in use will block
communication on the other.
Win-Situ 5 Software provides options for configuring analog/SDI-12 communications and Modbus communications on the
Tools
tab.
For additional information on Modbus and SDI-12 communications, including the SDI-12
commands and Modbus registers, see
www.in-situ.com/Technical_notes
and scroll to the
Communications and Software Technical Notes section.
Redundant Logging
The instrument is capable of internal logging while participating in a Modbus, SDI-12 or analog network. However, Win-Situ 5
Software cannot communicate with the instrument while it is transmitting Modbus, SDI-12, or 4-20 mA analog data, and the
instrument cannot receive or respond to Modbus, SDI-12, or 4-20 mA analog commands while connected to a PC serial port.
If the PLC or data logger loses data, the data that was logged internally on the instrument can be retrieved using Win-Situ 5
Software.
If the PLC or data logger experiences power loss, the instrument will continue to collect data using its internal batteries and
clock.
Содержание Level TROLL 400
Страница 51: ...1 800 446 7488 51 www in situ com See the online Help for more details ...
Страница 64: ...1 800 446 7488 64 www in situ com Analog 4 20 mA 2 Wire ...
Страница 65: ...1 800 446 7488 65 www in situ com SDI 12 3 Wire ...
Страница 66: ...1 800 446 7488 66 www in situ com Modbus Master ...
Страница 67: ...1 800 446 7488 67 www in situ com Modbus Master with RS232 Converter Required ...
Страница 78: ...1 800 446 7488 78 www in situ com Decontamination and Cleaning Form ...
Страница 81: ...1 800 446 7488 81 www in situ com Declarations of Conformity and Similarity Level TROLL 400 Instrument ...
Страница 82: ...1 800 446 7488 82 www in situ com Level TROLL 500 Instrument ...
Страница 83: ...1 800 446 7488 83 www in situ com Level TROLL 700 Instrument ...
Страница 84: ...1 800 446 7488 84 www in situ com TROLL Com Communication Device ...
Страница 85: ...1 800 446 7488 85 www in situ com BaroTROLL Instrument ...