- Two outputs for oscilloscope connection.
- Languages: Italian, English, French, German, Spanish.
- Improved use of the appliance based on a user-friendly system, thus allowing the user to operate quickly.

The standard equipment configuration includes: 
- Basic configuration of the appliance in a practical carrying case.
- Two 55kw piezometric probes with connecting cables.
- Calibrating cylinder, grease coupling agent
- NiMh 4,8V  2000mA rechargeable batteries pack with automatic low battery warning.
- Shockproof carrying case for the tester and the accessories.
- 230V external power pack and 12V 500m/A battery charger.

Weight: about 2 kg 

ATTENZIONE 

CORRECT CONNECTION: DO NOT connect the instrument TX output to the 
RX input of the instrument. 
TX and RX must be connected, by means of special cables, ONLY to the two 
probes.

WARNING 

BATTERIES CORRECT USE: remove the batteries’ pack from its slot and keep it 
separated if the equipment is not used for a long time. 
If  the  instruction  above  is  not  respected,  the  batteries’  pack  will  be  damaged 
quickly.
 

WARNING 

UNUSED  OF  THE  EQUIPMENT:  if  the  equipment  is  not  used  for more than 15 
days, the information and memorization of the date and time will be lost.  
At the first re-start, the operator has to set the correct date and time.
 

2.02  CASE CONTENTS 

  2 ultrasonic probes; 

  2 cables for probes, 3,5m each; 

  42.5µs sample 

  contact material 

  four “AA” rechargeable batteries 

  12V 500mA battery charger

 

WARNING

 

Avoid to give impacts to probes or to let them fall as there is the possibility to break the 
quartz crystals inside them. 

WARNING

 

This instrument, as almost all the electronic instruments, must not be stocked, when not 
used, in too cold or humid rooms, because otherwise when it is actuated, there is the 
possibility it does not work properly. Therefore before its utilization we suggest to keep it 
for some hours in a room having at least 20°C tempe rature and a low humidity 
percentage. 

2.03  OPERATOR INTERFACE 

Switching on 
In order to turn the appliance on, press the F1 key for about 5 seconds. 

www.impact-test.co.uk     

      www.impact-test.com      

 www.impact-testsets.co.uk 

Содержание CN842

Страница 1: ...User Guide User Guide www impact test com www impact test co uk www impact testsets co uk Ultrasonic Pulse Velocity Tester CN842 Impact Test Equipment Ltd www impact test co uk www impact test com www impact testsets co uk ...

Страница 2: ...est Equipment Ltd Building 21 Stevenston Ind Est Stevenston Ayrshire KA20 3LR T 01294 602626 F 01294 461168 E sales impact test co uk Test Equipment www impact test co uk Test Sieves Accessories www impact test com Test Sets to International Standards www impact testsets co uk ...

Страница 3: ...www impact test co uk www impact test com www impact testsets co uk 1 ...

Страница 4: ...ice are however present Such risks can be reduced by carefully following manual procedures using the suggested individual protection devices and respecting the legal and safety norms in force This manual includes Warning and Danger indications in relevant chapters These indications are shown with the words Danger or Warning in bold font and uppercase to make them highly visible WARNING indicates t...

Страница 5: ...re of the Instructions Manual 1 05 Composition of the Instructions Manual 1 06 Modifications and enclosures of the Instructions Manual 1 07 Constructor Identification 1 08 Machine Identification Data 1 09 EC Stamp 1 10 Usage 1 11 Operators 1 12 Storage 1 13 Transportation and Movement 1 14 Packaging removal Chapter 2 INSTRUCTIONS FOR USE 2 01 Main features 2 02 Case contents 2 03 Operator interfac...

Страница 6: ...conditions of danger immediately stop the machine should it be working irregularly and consult the authorised dealer s Sales Service department It is the Client s responsibility to verify at the time of installation and use that no conditions of use arise which are different to those indicated Refer to the Constructor when in doubt 1 11 OPERATORS WARNING DANGER The use transportation installation ...

Страница 7: ...ity of 70 WARNING The tester as almost all electronic appliances must not be put in too cold or humid environments when not used otherwise it could not work properly Therefore before its use it is recommended that the appliance is kept in environments with a low level of humidity and a minimum temperature of 20 1 13 TRANSPORTATION AND MOVEMENT WARNING In order to avoid irreparable machine damage m...

Страница 8: ...ime If the instruction above is not respected the batteries pack will be damaged quickly WARNING UNUSED OF THE EQUIPMENT if the equipment is not used for more than 15 days the information and memorization of the date and time will be lost At the first re start the operator has to set the correct date and time 2 02 CASE CONTENTS 2 ultrasonic probes 2 cables for probes 3 5m each 42 5µs sample contac...

Страница 9: ... of the tester are shown on a 128x64 pixel LCD display FUNCTIONS KEYBOARD 2 04 MAIN SCREEN By switching the appliance on the main screen becomes active and the operator can use the following functions in top down order Start of the acquisition Display of the configuration menu Display of the saved measures Download of the data through the mini USB port of the tester The main screen will be as the ...

Страница 10: ...quisition CONTINUOUS MODE The acquisition takes place in a mediated mode four measures are taken within a second checking that they fall within the range of the tester and that they are consistent with each other Otherwise the NOT VALID notice appears The operation is repeated three times checking the data consistency again Only in this case if the automatic save function is active data are saved ...

Страница 11: ... the main screen 2 06 CALIBRATION By selecting the appropriate entry of the configuration menu and by pressing the F2 key it is possible to perform a two points calibration in order to achieve the highest possible level of precision on the performed measures By starting the calibration the following screen appears START CALIBRATION EXIT This screen suggests the operator should bring the probes in ...

Страница 12: ... of the measure under consideration some values can be missing because of the deletion of one or more measures Tests identification in order to make the continuous mode tests more evident every single test is identified by a character which is gradually increased at the starting of any new test Time to fly it represents the elapsed time between the ultrasonic signal emission in the TX channel and ...

Страница 13: ...u To avoid interferences with the results it is advisable to keep the mini USB cable always disconnected from the tester during the measuring phase 2 09 CONNECTIONS The tester is provided with two sets of outwards connections In the upper part there are Mini USB connector for the communication with the host PC and for the data download 6 mm circular connector for the connection with the battery ch...

Страница 14: ...order to preserve the machine and its efficiency Avoid the use of solvents which damage paint and parts in synthetic material 3 03 AUTHORISED MAINTENANCE CENTRES For information on the nearest authorised help centre it is essential to contact the manufacturer Chapter 4 SPARE PARTS Accessories C370 08 Exponential couple of transmitting and receiving PROBES nominal frequency 55kHz C372 10 Couple of ...

Страница 15: ...ble Make the potential sources of danger harmless such as sharp or protruding parts Dismantle the machine divide it into similar parts and dispose of according to the standards in force Recycling notice for the disposal of electrical and electronical devices This symbol shown on the device or on the package and or the documentation suggests that the device should not be disposed together with othe...

Отзывы: