8
Meanwhile, some important information concerning the functioning of the appliance are displayed:
•
date and time;
•
battery level;
•
last measure;
•
acquisition mode (single/continuous)
•
automatic save
•
gain (250V/500V/750V/1000V)
By pressing the F1 key to start the sampling, the commands displayed in the following screen are enabled:
STOP OF THE MEASURING
The system reduces the number of commands, only the F1 key is enabled to stop the acquisition in case of continuous
mode.
SINGLE MODE
The acquisition takes place in a mediated mode; four measures are taken within a second, checking that they fall within
the range of the tester and that they are consistent with each other. Otherwise the notice “NOT VALID” appears. After
saving the data (if the automatic save function is active), the system completes the acquisition.
CONTINUOUS MODE
The acquisition takes place in a mediated mode; four measures are taken within a second, checking that they fall within
the range of the tester and that they are consistent with each other. Otherwise the “NOT VALID” notice appears. The
operation is repeated three times, checking the data consistency again. Only in this case, if the automatic save function
is active, data are saved (a short blink of “SAVE” sign indicates that data have been correctly saved).
2.05 CONFIGURATIONS
The configuration screen is the following:
MOVE THROUGH THE FIELDS
INCREASE / OK
DECREASE
EXIT
Press the F1 key many times to move through the different parameters of the tester, while the values can be
modified through the F2 (+) and F3 (-) keys.
•
MODE: the appliance includes two operating modes: a single mode, through which only a set of four
mediated readings is performed (thus creating only one value of measuring); a continuous mode, through