35
7.8 Offsetting the material temperature sensor
In the menu „Calibration“ and the window „Material
Temp Offset“, a zero point offset can be adjusted for the
material temperatur sensor which is installed inside the
SONO probe. In this example a temperature deviation of
+5° C is produced by inside self-warming of the SONO
probe. The correction value -5 can be setted in the
Coeff0 window.
The example shows the parameters for displaying the
temperature in the unit: Degree Fahrenheit.