Calibration 

Curve

Recommended for 

grain type

Bulk density 

of grain type

Application

Cal1

Maize, without TC 

(TC = Temperature 

Compensation)

0,75

Installation at the discharge hopper. The outfeed is 

batch by batch and it is not secured, that the 1-rod 

probe is continually covered with grain. 

Cal2

Maize with TC

0,75

Installation inside a silo where the 1-rod probe is 

continually covered with grain. 

Cal3

Wheat without TC 

0,75

Pre-installed like Cal1

Cal4

Wheat with TC

0,75

Pre-installed like Cal2

Cal5

Rye without TC

0,72

Pre-installed like Cal1

Cal6

Rye with TC

0,72

Pre-installed like Cal2

Cal7

Barley without TC 

0,63

Special calibration curve

Cal8

Barley with TC

0,63

Special calibration curve

Cal9

Rape and oilseeds 

without TC

0.65

No temperature compensation necessary!

Cal10

Sunflower seeds wit-

hout TC

0,30

No temperature compensation necessary!

Cal11

Soya without TC

0,65

Special calibration curve

Cal12

Soya with TC

0,65

Special calibration curve

Cal13
Cal14 

(pre-setted 

after delivery)

Corn ear of maize

0,5

Special calibration curve

Cal15

1/10 tp

Radar time and reference calibration for test

5.4  Selection and application of the reference method

In order to adjust the TRIME-GW Line for precise absolute measurements at the discharge, an off-line  measurement 

method must be available to serve as a reference. It must provide a high degree of absolute precision and function 

with large sample volumes.
The TRIME-GW Line measures the average value continuously over a large measurement volume. In moving grain, 

the results of multiple measurement volumes can be accumulated in respect to the chosen avereging time. It 

therefore requires a lot of time and effort to verify this very representative value with a reference instrument 

mostly using much more smallersample quantities. The most suitable method for determining the exact moisture 

of the grain is to use a drying oven. The sample volume is of importance and should be at least 0.5 litres.
For sample extraction and making reference measurements please take the following best practice points into 

account: 

Содержание TRIME-GW Series

Страница 1: ...Manual TRIME GW Line More information www imko de Moisture Sensor Experts...

Страница 2: ...oisture probe You may contact IMKO directly to after exploring your local contact We look forward to helping you For warranty claims please contact your local dealer distributor or service center The...

Страница 3: ...allation on a steel girder with HEA profile 12 4 3 Carrier plate dimensions for 120mm steel girder 13 4 4 Installation of two carrier plates very close to each other 15 5 Initial operation and adjustm...

Страница 4: ...ue outputs of the SONO probe 31 7 4 Configuration of Measure Mode 32 7 5 Setting the precision of a single value measurement cycle 33 7 6 Test run in the respective Measurement Mode 33 7 7 Measure Run...

Страница 5: ...l capacitive or microwave measuring methods the TRIME technology Time Domain Reflectometry with Intelligent Micromodule Elements offers precise measurement results which means more reliability at the...

Страница 6: ...te the electronic separately from the 1 rod probe TRIME GW Line offers two possibilities for temperature compensation A Temperature compensation of the internal TRIME electronic Despite the TRIME GW L...

Страница 7: ...ME GW Line can be requested via Contact us on www imko de 2 6 The IMP Bus as a user friendly network system With external power supply on site for the SONO probes a simple 2 wire cable can be used for...

Страница 8: ...ways to configure and adjust a SONO probe A Online Configuration via SONO VIEW Withthestand alonedeviceSONO VIEWitispossibletoconfigureSONO probesonlinetoanappropriate operating mode without the need...

Страница 9: ...propriate operating mode like cyclic measurement averaging filtering cumulating and other powerful operating parameters Furthermore it is possible to select a calibration curve inside a SONO probe wit...

Страница 10: ...at the sensor The plant must be properly grounded Transparent Transparent Assignment of the 10 pole MIL Plug and sensor cable connections 3 4 Analogue Output 0 10V with a Shunt Resistor With the help...

Страница 11: ...11 4 TRIME GW Line Installation...

Страница 12: ...HEA profile TRIME GW Line is delivered with a carrier plate as U profile This carrier plate can be mounted directly on a steel girder which can be installed in silos or drying chambers IMKO offers th...

Страница 13: ...13 4 3 Carrier plate dimensions for 120mm steel girder...

Страница 14: ...14 Carrier plate dimensions for 4 steel girder...

Страница 15: ...stallation of two carrier plates very close to each other Here it is very important that the two long rods have sufficient distance to one another This can be achieved with a spacer between the carrie...

Страница 16: ...ired calibration curve Cal1 Cal15 and with using the button Set Active Calib The finally desired and possibly altered calibration curve Cal1 15 which is activated after switching on the probes power s...

Страница 17: ...17 5 3 Calibration Curves Cal1 to Cal15 Calibration functions with and without temperature compensation...

Страница 18: ...5 Special calibration curve Cal13 Cal14 pre setted after delivery Corn ear of maize 0 5 Special calibration curve Cal15 1 10 tp Radar time and reference calibration for test 5 4 Selection and applicat...

Страница 19: ...ifferent point than in the direct vicinity of the probe are not representative Several samples must be taken directly at the probe and averaged At the start of trial operation the suitable calibration...

Страница 20: ...ied with the calibration coefficients see Chapter Creating a linear calibration curve The desired and possibly altered calibration curve Cal1 15 which is activated after switching on the probes power...

Страница 21: ...But it would be also possible to use a higher point P1 e g 190 7 where a tpAve of 190ps corresponds to a moisture content of 7 The gravimetric moisture content of the material e g 7 has to be determin...

Страница 22: ...dDcanbealsoenteredwithoutaPCwiththemoduleSONO VIEW seemanualSONO VIEW 5 11 Calculation for a non linear calibration curve SONO probes can work also with non linear calibration curves with polynomials...

Страница 23: ...rage value can be set with this parameter CC CH CK Setting of the time for calculation of the trend or expectation value for the Boost Offset function Filter Upper Limit Offset Standard Setting 25 Set...

Страница 24: ...nt Standard Setting 2 values Setting Range 0 10 Unit Measurement Values with 3 single values per second CA CF CK inactive CC CH Number of discarded poor measurement values after the start of a new bat...

Страница 25: ...ly partially filled with material for a longer time i e the material temporarily no longer completely fills the probe measuring field During these periods Lower Limit Keep Time the probe would establi...

Страница 26: ...atch process starts So the PLC has time enough to read in the freezed moisture value of the batch For applications without a PLC the freezed signal of the TRIME GW Line can be used for displaying the...

Страница 27: ...value is freezed at the analogue output until the No Material Delay time of e g 5 seconds has expired and a new batch process starts Start of second batch process After the period of No Material Dela...

Страница 28: ...USB to the USB Port of the PC and the installation will be accomplished automatically Install Software SONOConfig SetUp msi from USB Stick Connection of the SONO probe to the SM USB with 4 wires for...

Страница 29: ...there is no double occupancy of the interfaces In case of a conflict when the serial port or the USB SM has been found in a higher COM port the COM port number can be adjusted manually By double click...

Страница 30: ...isplayed on the screen In the menu Bus and the window Configuration the PC can be configured to an available COMx port with the Baudrate of 9600 Baud 7 1 Scan of connected SONO probes on the serial in...

Страница 31: ...can be selec ted to IMP Bus or RS485 Due to very robust behavior it is recommended to select the IMP Bus 7 3 Set analogue outputs of the SONO probe In the menu Config and the window Analog Output the...

Страница 32: ...ent measurement modes e g for averaging values from several single measurement values for performing a filtering or performing other functions see chapter Measurement mode configuration in this manual...

Страница 33: ...Quick Precise Mode C e g CS CH CF etc Quick and precise TDR pulse search and precise measurement within 350ms Similar to Quick but with a little longer reaction time in process measurements Mode A Mod...

Страница 34: ...for the averaging tpAve Radar time average which corresponds to the respective moisture value By clicking Save the recorded data is saved in a text file in the following path SONO CONFIG exe Pfad MD D...

Страница 35: ...adjusted for the material temperatur sensor which is installed inside the SONO probe In this example a temperature deviation of 5 C is produced by inside self warming of the SONO probe The correction...

Страница 36: ...parameter TempComp can be adjusted for higher temperature ranges up to 120 C for special high temperature version to values up to TempComp 0 75 But it is to consider that it is neces sary to make a B...

Страница 37: ...epends on the used material Repeatability 0 2 b w depends on the measured material Measurement transformer tempera ture range 10 60 C extended range on request Probe temperature range 0 60 C Measuring...

Страница 38: ...out of operation and secured by means of marking before further commissioning In case of doubt the sensor or the measuring system must be sent to the manufacturer or his contractual partner for repai...

Страница 39: ...in following case if you do not use the sensor or the measuring system for an extended period of time before cleaning unpacking or changing the sensor or the measuring system if you are working inside...

Страница 40: ...Contact IMKO Micromodultechnik GmbH Am Reutgraben 2 76275 Ettlingen Germany Tel 49 7243 5921 0 Fax 49 7243 5921 40 info imko de Manual TRIME GW Line 08 2021 www imko de...

Отзывы: