You must also consider the degradation of the LED when saving self-defined presets. If you 
save a preset with LED channels that uses its maximum intensity, the possibility exists that this 
intensity cannot be reached after the burn-in time or the long-time degradation of the LED. In 
this case, you will get a warning message from the iQ-LED control software. 

During the first 500-600 working hours, we recommend performing a spectral calibration every 
50 operating hours. 

After the first 500-600 operating hours, a calibration of every 150 working hours suffices. 

Other factors that indicate the need for a spectral calibration: unsatisfactory illuminant 
generation, the aberration of the intensity values, or a spectral curve that does not fit with the 
predefined standard illuminants of the corresponding preset.  

  the spectrometer works correctly 

  the spectrometer settings are correct 

  all LED channels work correctly 

  the dark measurement is correct 

  your measurement environment is correct 

  your ambient temperature is correct 

 

How to perform the spectral calibration is described in the iQ-LED control software manual. 

4.4  Low intensity use 

When using your system with very low intensity, the spectral measurement values will start to 
fluctuate. The lower the intensity, the higher the fluctuation. The generated light is still stable up 
to a certain point. The fluctuation of the values is caused by the noise of the spectral 
measurement of the internal spectrometer. The light intensity will continue to get lower when the 
influence of the noise continues to get higher. When using standard illuminants with an intensity 
lower than 25 lux, it will no longer be possible to get a correct value.   

 

5  ADDITIONAL INFORMATION 

5.1  Maintenance 

The spectrometer requires a recalibration once a year, regardless of the operating hours. If a 
spectrometer calibration is necessary, please contact Image Engineering. We will provide you 
with removal instructions for the spectrometer. The spectrometer then has to be sent to our 
office with a notation for calibration. Use appropriate packaging to prevent the spectrometer 
from being damaged during shipping.  

A temporary replacement spectrometer can be provided to bridge the time gap while your 
spectrometer is recalibrated. Please contact us for conditions and procedure.  

 

 

 

Содержание CAL2

Страница 1: ...Image Engineering GmbH Co KG Im Gleisdreieck 5 50169 Kerpen Horrem Germany T 49 2273 99991 0 F 49 2273 99991 10 www image engineering com CAL2 User Manual 18 December 2018 ...

Страница 2: ...RUCTIONS HARDWARE 5 3 1 Overview display and ports 5 3 2 Connecting the hardware 6 3 3 Camera positioning 7 4 OPERATING INSTRUCTIONS SOFTWARE 7 4 1 Requirements 7 4 2 Software installation 7 4 3 Starting the system 7 4 3 1 Spectrometer settings 7 4 3 2 Spectrometer calibration 8 4 3 3 iQ LED calibration 8 4 4 Low intensity use 9 5 ADDITIONAL INFORMATION 9 5 1 Maintenance 9 5 2 Care instructions 10...

Страница 3: ...empered environment without any interfering light The optimal ambient temperature range is 22 to 26 degrees Celsius The maximum ambient temperature range is 18 to 28 degree Celsius The optimal system temperature range displayed in the software user interface is between 35 and 50 degrees Celsius The system has internal temperature management if there is any error regarding the internal temperature ...

Страница 4: ...tions from the Image Engineering support team or when connected to the power supply 2 GETTING STARTED 2 1 Scope of delivery CAL2 calibration device spectrometer power cord USB cable control software calibration protocol 2 plates for connecting the spectrometer Optional equipment iQ Analyzer software shading module This module includes a special chart layout file that allows analysis of images with...

Страница 5: ... storage of the illuminants with the play and stop button you can start and stop a saved light sequence with different illuminants it is possible to save one sequence on the device with the power button you can turn on and off the light there are three pre stored illuminants on your device the intensity of each illuminant is shown in the acceptance protocol of your device o 1 illuminant A default ...

Страница 6: ...d to synchronize your test setup For example with an iQ Trigger See 2 1 optional equipment 3 2 Connecting the hardware 1 Connect the power cord to the power supply on the back of CAL2 2 Connect the USB cable to the CAL2 and your PC 3 Turn on the CAL2 the power switch is located beside the power supply 4 Connect the spectrometer to the PC using the USB cable The system will install the spectrometer...

Страница 7: ...art the iQ LED software by clicking the iQ LED exe or the iQ LED icon on your desktop Follow the iQ LED software manual to control the CAL2 NOTE The iQ LED devices can only operate with high precision when setup and calibration are performed correctly Consult the iQ LED software manual for a comprehensive description and read it carefully 4 3 1 Spectrometer settings When the spectrometer is connec...

Страница 8: ...lly NIST traceable calibrated We recommend recalibrating the spectrometer once a year regardless of the operating hours If a spectrometer calibration is required please contact Image Engineering Note Before removing the spectrometer measure and note the lux value of a predefined standard illuminant After the spectrometer is calibrated and has been reinstalled to your system perform a spectral cali...

Страница 9: ...rrect How to perform the spectral calibration is described in the iQ LED control software manual 4 4 Low intensity use When using your system with very low intensity the spectral measurement values will start to fluctuate The lower the intensity the higher the fluctuation The generated light is still stable up to a certain point The fluctuation of the values is caused by the noise of the spectral ...

Страница 10: ... calibration is invalid and the spectrometer has to be recalibrated Only store and transport the CAL2 in the delivered hard case 5 3 Disposal instructions After the service life of CAL2 it must be disposed of properly Electrical and electromechanical components are included in CAL2 Observe your national regulations Make sure that CAL2 cannot be used by third parties after disposing of it Contact I...

Страница 11: ...ical data sheet Image Engineering CAL2 Seite 11 von 11 6 TECHNICAL DATA SHEET See annex for the technical data sheet It can also be downloaded from the website of Image Engineering www image engineering com ...

Отзывы: