C A L I B R A T I O N A N D T R O U B L E S H O O T I N G
Remote Calibration of the LDX-32420 Current Source
78
LDX-32420
C H A P T E R
6
Remote Calibration of the LDX-32420 Current Source
The LDX-32420 Current Sources can be calibrated remotely. All of the required
calibration commands are listed in Table 4.1 in Chapter 4. The procedures below
are general guidelines for writing a program to calibrate your instrument.
Note:
The operation complete flag (bit 0 of the Standard Event Status Register) may be
used to trigger a service request. This type of interrupt is enabled by setting bit 0 of the
Service Request Enable register (via the
*ESE
command) and bit 5 of the Service Request
Enable register (via the
*SRE
command). Service request (SRQ) handling depends on your
GPIB hardware. Refer to your GPIB user's manual for details.
After a calibration value is entered, the
*OPC?
query may be used to determine when the
calibration sequence is done. However, the
*OPC
,
*WAI
command, or
*OPC?
query should
not be issued until after the expected calibration value is entered, or the system will "hang".
This happens because the LDX-32420 Current Source will wait indefinitely for an input, yet
not allow any input until the calibration is finished.
Current Source Calibration
There are three calibration adjustments required for the LDX-32420 Current
Source. They are calibration of the constant current source for both bandwidths
and ranges, calibration of the laser voltage measurement, and calibration of the
constant light power (I
PD
) feedback circuits.
The LDX-32420 Current Source implements a two-point calibration for the current
source. Two currents (approximately 20% and 80% of FS) are applied to a test
load, and the resulting actual currents are fed back (by the user) to the current
source. The laser current source calibration program uses the two sets of data to
calculate new calibration constants. These calibration constants are used to set
the actual current of the current source.
The following procedure is for calibrating each range of the current source. Start
with the low range. After calibration is complete in this range, switch to the high
range and repeat the procedure.
Configure the current source in the following manner for the laser current source
calibration:
Function
State
COMMAND
LASER MODE
I (constant current low bandwidth)
LAS:MODE:ILBW
RANGE
Low (lowest per instrument model)
LAS:RAN x
LIMIT
90% of FS (full scale of range)
LAS:LIM:Ix
SETPOINT
80% of FS
LAS:LDI
Содержание LDX-32420
Страница 3: ......
Страница 9: ...L I S T O F FIG U R E S vi LDX 32420...
Страница 11: ...L I S T O F TA B L E S viii LDX 32420...
Страница 17: ...WA R R A N T Y xiv LDX 32420...
Страница 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Страница 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Страница 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...