R E M O T E O P E R A T I O N
Error Messages
01_10
LDX-32420
39
C H A P T E R
3
Error Code
Explanation
E-205
<PROGRAM DATA> is not a Boolean value or word
E-207
<PROGRAM DATA> will not convert to an unsigned 16-bit value
E-209
<PROGRAM DATA> will not convert to an unsigned 32-bit value
E-210
<PROGRAM DATA> will not convert to a floating point value
E-211
<PROGRAM DATA> will not convert to a character value
E-213
<PROGRAM DATA> is incorrect block data length
E-214
<PROGRAM DATA> length exceeds maximum
E-302
Query error; device was addressed to talk, but GPIB controller failed to read all of the
<RESPONSE MESSAGE>
E-501
LASER interlock disabled output
E-503
LASER open circuit disabled output
E-504
LASER current limit disabled output
E-505
LASER voltage limit disabled output
E-507
LASER power limit disabled output
E-510
LASER out of tolerance disabled output
E-511
LASER control error disabled output
E-512
Analog section status is all 1’s or all 0’s (power down)
E-513
Serial EEPROM checksum error
E-515
Laser output must be off to change ranges
E-516
Incorrect configuration for calibration sequence to start
E-519
Setting a measurement is only valid during the calibration phase for that measurement;
user has tried to calibrate a measurement without first entering the required calibration
mode.
Содержание LDX-32420
Страница 3: ......
Страница 9: ...L I S T O F FIG U R E S vi LDX 32420...
Страница 11: ...L I S T O F TA B L E S viii LDX 32420...
Страница 17: ...WA R R A N T Y xiv LDX 32420...
Страница 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Страница 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Страница 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...