The LASer:CAL:LDV command is used to enter the LASER voltage measurement
calibration mode.
The LASer:CAL:LDV? query is used to determine that the LDX-32420 is ready for a
value to be entered during the calibration cycle of the LASer:CAL:LDV mode.
Notes
After LASer:CAL:LDV is issued, the LDX-32420 will allow calibration of the laser voltage
measurement. This procedure is outlined in Chapter 6.
Examples
"Las:CAL:Ldv" - action: the LDX-32420 enters calibration mode for LASER voltage.
"LASer:CAL:LDV?" - response: 1, means the LDX-32420 is ready for the user to enter a
voltage value via the LASer:LDV command.
The LASer:CAL:MDI command is used to enter the LASER photodiode current
calibration mode.
The LASer:CAL:MDI? query is used to determine that the LDX-32420 is ready for a value
to be entered during the calibration cycle of the LASer:CAL:MDI mode.
Notes
After LASer:CAL:MDI is issued, the LDX-32420 will automatically enter the LASER
photodiode current calibration mode. When the LDX-32420 is ready, the user should
enter the measured value. This procedure is outlined in Chapter 6.
If the LASER output is not ON or if the P mode is not selected, error code E-516 will be
generated.
Examples
"Las:CAL:MDI" - action: the LDX-32420 enters the LASER photodiode current
calibration mode.
LAS:CAL:MDI?" - response: 1, means the LDX-32420 is ready for the user to enter a
photodiode current value via the LASer:MDI command
LASer:CAL:LDV
LASer:CAL:LDV?
F
RONT
P
ANEL
R
EMOTE
LASer:CAL:MDI
LASer:CAL:MDI?
F
RONT
P
ANEL
R
EMOTE
Содержание LDX-32420
Страница 3: ......
Страница 9: ...L I S T O F FIG U R E S vi LDX 32420...
Страница 11: ...L I S T O F TA B L E S viii LDX 32420...
Страница 17: ...WA R R A N T Y xiv LDX 32420...
Страница 41: ...O P E R A T I O N Operating a Laser in Constant Power P Mode 24 LDX 32420 C H A P T E R 2...
Страница 57: ...RE M O T E O PE R A T I ON Error Messages 40 LDX 32420 C H A P T E R 3...
Страница 103: ...C A L IB R A TI O N A N D TR OU BL E S HO O T IN G Error Messages 86 LDX 32420 C H A P T E R 6...