Description 

3 Description 

3.1 Design 

A filter holder for a diaphragm filter is integrated between drip pan 

(1)

 and diaphragm pump 

(2)

. A diaphragm filter up to a filter fineness of 0.22 µm can be inserted into this holder by the 

user as required for the work to be done. According to the filter fineness used, it is possible 
to achieve different types of filtration, enabling, among other uses, the

 handling of bacteria 

and viruses

.

 

The sealing 

(3)

 at the head of the drip pan consists of a formed gasket. 

The connections to the diaphragm pump are established by using quick-fitting couplings 

(4)

which are provided with a two-side stop valve each on both sides. 
The chemical-resistant diaphragm pump used has a long useful life and at the device switch 
one switches on. If necessary switching on can take place also via foot switch 

(5)

 (accesso-

ries optional). 
Drip pan 

(1)

 and handpiece (pipette holder) 

(6) 

are connected to one another using a hose. 

The built-in hose coupling 

(7)

, which seals both sides, can be used to shut off the connec-

tion. 
The user can insert suitable pipetting tips into the receptacle of the handpiece (pipette 
holder). 
 
 

(7)

(5)

(2)

 
 
 
 

(3) 

 
 
 
 

(1) 

 
 

(4) 

 
 
 
 
 

(6) 

 
 
 
 
 
 
 
 
 

Fig. 1   Safety Aspiration-System “biovac 106” 

8 112012 

Содержание biovac 106

Страница 1: ...Operation Manual Safety Aspiration System Type biovac 106 112012 2008 04 04...

Страница 2: ...owed without the prior written permission of ILMVAC GmbH All rights under the copyright laws are expressly reserved by ILMVAC GmbH We reserve the right to make changes and amendments ILMVAC GmbH Am Vo...

Страница 3: ...Accessories 9 4 Technical Data 10 4 1 Intake Pressure Pumping Speed Diagram 10 5 Installation and Operation 11 5 1 Unpacking 11 5 2 Installation and Connection 11 5 2 1 Schematic representation of the...

Страница 4: ...ion System must have the technical competence required to perform the work that has to be done The user must authorize the operating personnel to do the work that has to be done The personnel must hav...

Страница 5: ...tions EN 61010 1 Safety for laboratory devices EN 50081 1 2 Electromagnetic compatibility EMC Generic specification Interference resistance for residential business and industrial areas and small busi...

Страница 6: ...ase of portable devices The local regulations apply in foreign countries Please note the following when connecting to the electrical power supply system The electrical power supply system must have a...

Страница 7: ...gases Prevent such substances from leaking out of the appliance or pump Treat these substances according to the applicable environmental protection regulations The diaphragm pump control valves and h...

Страница 8: ...ket The connections to the diaphragm pump are established by using quick fitting couplings 4 which are provided with a two side stop valve each on both sides The chemical resistant diaphragm pump used...

Страница 9: ...s soon as the maximum level is reached 3 3 Areas of Application The ILMVAC biovac 106 Safety Aspiration System is intended for the safe and precise extraction of liquid and non explosive excess fluids...

Страница 10: ...ature C 40 Bearing maintenance free Reference surface sound pressure level DIN 45635 part 13 dB A 45 Voltage AC asynchronous motor V Hz 115 230 Frequency Hz 50 60 Power W 0 45 Fine fuse A F 3 15 Opera...

Страница 11: ...claim under warranty the device must be returned in packaging that is suitable for protecting it during transport 5 2 Installation and Connection 1 Set the safety aspiration system on a flat and horiz...

Страница 12: ...ower supply either 115 or 220 V 10 Set mode selector switch to Hand or Foot The safety aspiration system is switched on and off at the operating switch illuminated rocker switch Press the foot switch...

Страница 13: ...tection may be used 5 5 Scrap Disposal CAUTION The safety aspiration systems must be disposed of in accordance with the 2002 96 EC guideline and the specific national regulations Contaminated pump sy...

Страница 14: ...materials and clean the pump thoroughly before handing it over to third parties The operator may perform maintenance work to the extent indicated below 6 2 Maintenance of the aspiration system 6 2 1...

Страница 15: ...he membrane filter at regular defined intervals Test performance of the quick fitting couplings with shut off valves WARNING Only perform the work that is described here and that which is permitted to...

Страница 16: ...Connected apparatus leaks connecting elements leak Identify and seal the leak replace the seals and hoses if necessary Diaphragm pump leaks pipes or hoses Check the hose connections between the pump h...

Страница 17: ...quantity serial number and order number CAUTION Ilmvac is not liable for any damage caused by the installation of any parts not sup plied by the manufacturer 8 1 Maintenance kit Designation Order no...

Страница 18: ...Spare Parts Overview 8 2 Spare parts view Fig 5 Exploded view biovac 106 18 112012...

Страница 19: ...ocker switch 16 4 A green 1 825186 1 29 Voltage voter 115 230 V 1 825187 30 Socket for non heating apparatus 1 825274 5 31 Casing foot 5 829112 32 Rubber plate 1 112547 02 33 Guard ring 1 112561 34 Co...

Отзывы: