Maintenance and Servicing
6 Maintenance
and
Servicing
6.1 General
Requirements
Repairs of the ILMVAC biovac 106 safety aspiration-system may only be performed by the
manufacturer or authorized workshops. The prerequisites are a complete and factually cor-
rect damage report, and a clean and, if necessary, a decontaminate device.
Send in defective devices for repair either to the manufacturer or to an authorized workshop.
The information about the contamination or thorough cleaning are legally binding parts of the
contract.
WARNING !
When repairing contaminated units, be sure to observe the applicable user specifica-
tions regarding decontamination as required.
Provide full information about the type of contamination and the used materials and
clean the pump thoroughly before handing it over to third parties.
The operator may perform maintenance work to the extent indicated below:
6.2
Maintenance of the aspiration-system
6.2.1 Disassembly
•
Disconnect the power supply and ensure that it cannot be switched on again.
•
Disconnect the hose connections at the couplings.
•
To empty, clean and decontaminate the collecting tank, lift it up out of the holder of the
aspiration system.
•
Unscrew from below the four inner screws on the device.
•
Unscrew the suction and exhaust connections.
•
Lift the hood off.
You can now start maintenance work as described in the appendix to the
"Diaphragm
Pump"
operating instructions.
WARNING !
Wear protective gloves or protective clothing if necessary !
Do not clean with compressed air !
6.2.2 Assembly
Assembly takes place in reverse sequence. Use care when carrying out these activities, and
check for the proper fit of the couplings.
14 112012