4
Immunity test
IEC 60601 test level
compliance level
Electromagnetic environment
guidance
Electrostatic
㻌
discharge(ESD)
IEC61000-4-2
±6kV contact
±8kV air
±6kV contact
㻌
±8kV air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
±2kV for power
supply lines
±1kV for input/output
lines
±2kV for power
supply lines
±1kV for
input/output lines
Main power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC61000-4-5
±1kV differential
mode
±2kV common mode
±1kV differential
mode
±2kV common mode
Main power quality should be
that of a typical commercial or
hospital environment
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC61000-4-11
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
<5% Ut
(>95% dip In Ut)
for 0.5 cycle
40% Ut
(60% dip In Ut)
for 5 cycle
70% Ut
(30% dip In Ut)
for 25 cycle
<5% Ut
(>95% dip In Ut)
for 5 cycle
Main power quality should be
that of a typical commercial or
hospital environment. If the user
of the MODEL MKC-750UHD
requires continued operation
during power mains
interruptions, It is recommended
that the MODEL MKC-750UHD
be powered from an
uninterruptible power supply or
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at characteristic of a
typical location In a typical
commercial or hospital
environment.
NOTE It is the a.c.mains voltage prior to application of the test level.
Guidance and manufacturer's declaration - electromagnetic emissions
The Model MKC-750UHD is intended for use in the electromagnetic environment specified below. The
customer or the user of the Model MKC-750UHD should assure that it is used in such an environment.
Emissions test
Compliance Electromagnetic
RF emissions
㻌
CISPR 11 Group 1
The Model MKC-750UHD uses RF energy only for its
internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
㻌
CISPR 11 Class B
The Model MKC-750UHD is suitable for use in all
establishments, including domestic establishments and
those directry connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC61000-3-2
Class A
Voltage
fluctuations/flicker
emissions IEC61000-3-3
Complies
Содержание MKC-750UHD
Страница 1: ...MKC 750UHD Digital Process Compact 3CMOS Color Camera Operation Manual...
Страница 8: ......
Страница 22: ...When the automatic adjustment is complete the light of the auto white button will go out automatically...
Страница 34: ...OFF Disable an adjustment ON Enable an adjustment...
Страница 60: ...8 Appearances Camera Head...
Страница 61: ...CCU 270 210 8...
Страница 62: ...MEMO...