3
Guidance and manufacturer's declaration - electromagnetic emissions
The Model MKC-210HD is intended for use in the electromagnetic environment specifi ed below. The customer or the user
of the Model MKC-210HD should assure that it is used in such an environment.
Item
Test Procedure
Limits
Result
AC Mains Terminal Continuous
Disturbance Voltage
(0.15 – 30MHz)
CISPR 11
Group1,
Class A
Complied
Radiated Electric Field Strength
(30 – 1000MHz)
CISPR 11
Group1,
Class A
Complied
AC Mains Terminal
Harmonics
IEC61000-3-2
Class A
Complied
AC mains Terminal
Voltage Fluctuation
IEC61000-3-3
Clause 5
Complied
Immunity test
Specifi cation
Actual Performance
Result
Electrostatic discharge
IEC61000-4-2
±2, ±4,±6kV (contact)
±2, ±4,±8kV (air)
±2, ±4,±6kV (indirect contact)
±2, ±4,±6kV (direct contact)
±2, ±4,±8kV (air)
Complied
Radio-frequency
electromagnetic fi eld
IEC61000-4-3
80MHz - 2.5GHz
3 V/m (1kHz, AM 80%)
80MHz - 2.5GHz
3.6V/m(unmodulated, r.m.s)
(1kHz, AM 80%)
Complied
Electrical fast
transient/burst
IEC61000-4-4
AC power ports
±2kV (5/50ns)
±2kV (5/50ns,5kHz and 100kHz)
Complied
Signal ports
±1kV (5/50ns)
±1kV (5/50ns,5kHz and 100kHz)
Complied
Surge
IEC61000-4-5
AC power ports
±2kV (line-earth)
±1kV (line-line)
(90°/180°/270°)
±0.5, ±1, ±2kV (line-earth)
±0.5, ±1kV (line-line)
(90°/180°/270°)
Complied
Ratio-frequency
Conducted disturbances
IEC61000-4-6
AC power ports
0.15 - 80MHz,
3V (1kHz, AM 80%)
0.15 – 80MHz
3.15V (unmodulated, r.m.s)
(1kHz, AM 80%)
Complied
Signal ports
0.15 - 80MHz,
3V (1kHz, AM 80%)
0.15 – 80MHz
3.15V (unmodulated, r.m.s)
(1kHz, AM 80%)
Complied
Power frequency
magnetic fi eld
IEC61000-4-8
3 A/m,50Hz and 60Hz
3.15 A/m, 50Hz and 60Hz
Complied
Voltage dips
and short interruptions
IEC61000-4-11
>95% reduction, 0.5periods
60% reduction, 5periods
30% reduction, 25periods
(Changes to occur at 0degree
crossover Point of the voltage
waveform)
>95% reduction / 0.5periods
(0°/180°)
60% reduction / 5periods (0°)
30% reduction / 25periods( 0°)
Complied
>95% reduction, 5s
(Changes to occur at 0degree
crossover point of the voltage
waveform)
>95% reduction / 5ns(0°)
Complied