Unpacking
• Unpacking:
- Unpack the device carefully.
- Any damage should be notified immediately to the shiping
agent (post, rail or logistic company).
• Delivery scope:
-
STARVISC 200-2.5 control station
with a
WiCo
- Extension arm
- Hexagonal socket screw
- Hexagon socket offset screw key
- Chuck key
- Protective cover
- Pt 1000 temperature sensor
- OS 1.0 power supply unit
- USB 2.0 cable (micro A – micro B)
- USB 2.0 cable (A – micro B)
- User guide
- Warranty card
Correct use
• Use:
For mixing/stirring liquids with low and high viscosities by vari-
ous stirring tools.
Intended use: Stand device (chuck points down).
• Range of use (indoor use only):
- Laboratories
- Schools
- Pharmacies
- Universities
• Wireless remote control:
Before using the wireless link between the
WiCo
and the labo-
ratory device, first check whether your region is included in the
radio communications approval for the device. If it is not, remote
control can also be performed using a USB cable.
The device is suitable for use in all areas (EMC class A and B).
Note:
close to the magnetic material
may affect the accuracy of the torque
measuring.
The safety of the user cannot be guaranteed:
- if the device is operated with accessories that are not supplied or
recommended by the manufacturer.
- if the device is operated improperly or contrary to the manufac-
ture’s specifications.
- if the device or the printed circuit board are modified by third parties.
Drive
The rotary knob (B, see Fig. 1) of the
WiCo
allows the speed of the stirrer to be adjusted over the entire speed range.
OS 1.0 power supply unit
(for
WiCo
):
Adapter
England
Adapter
USA, China
Adapter
Australia
Adapter
Europe, Switzerland
Motor protection
The stirring device is suitable for continuous operation. The motor
current is limited electronically. The device has an anti-stall and
anti-overload system.
If a fault occurs, a safety circuit immediately switches off the mo-
tor permanently through a relay in the power board. A fault occurs
if the safe functioning of the device is compromised.
Speed – normal operation
Speed - regulated (no variation in speed):
The speed is monitored and regulated by processor controlled.
The target speed is constantly compared with the actual rota-
tion speed value of the output shaft and variations corrected. This
guarantees a constant speed even if the viscosity of the substance
being stirred changes.
Fluctuations in mains voltage within the permitted tolerance range
have no effect on the quality of regulation and constancy of speed.
The speed is set using the front knob (B, see Fig. 1). During normal
operations the speed value on the display (C, see Fig. 1) corresponds
to the speed of the output shaft in revolutions per minute (rpm).
NOTICE