19
Safety:
Time Out:
In the “Time Out“ menu, you can set a time limit. This time limit
applies if there is a communication failure between the
station
and the
WiCo
or the communication range is exceeded. The
station
will continue to run at the set speed until the set time has expired.
After that the stirrer (station) will continue to run at the set safe
speed (see the "Safe speed" menu).
Note:
The initial setting for the time limit is 30 seconds. You can
set a value of up to 60 minutes for this time limit.
If the intermittent mode is activated, the
station
immediately switches to run at the set safe
speed, or the set speed if that is less than the safe speed.
Safe Speed:
In the “Safe Speed“ menu, you can specify a speed that is appropri-
ate and safe for the stirring task. The safe speed applies if there is a
communication failure between the
station
and the
WiCo
or the
communication range is exceeded.
Note:
The initial setting of the safe speed is 100 rpm is imple-
mented after expiry of the time limit (see "Time out").
Password:
In the “Password“ menu, you can protect the WiCo settings using a
password (factory setting: 000).
Settings:
Languages:
The “Languages“ option allows you to select the desired lan-
guage by turning and pressing the rotary/push knob (B). A tick
indicates the language that is set for the system.
Units:
The ”Units“ option allows you to select the desired unit for dis-
playing the temperature. The choice is between ”°C” or ”°F” and
is made by turning and pressing the rotary/push knob (B). A tick
indicates the unit that is set for the system.
Display:
The ”Display“ option allows you to change the background color
and brightness of the working screen.
Sound:
The ”Sound” option allows you to activate/deactivate the key-
press sound and to set the volume.
Factory Settings:
Select the ”Factory Settings” option by turning and pressing the
rotary/push knob. The system requests confirmation to recreate
the factory settings. Pressing the ”OK” button resets all the sys-
tem settings to the original standard values set at dispatch from
the factory (see "Menu structure").
Communication:
The ”Device name” menu option allows you to edit the de-
vice name. This can relatively identify the paired station.
The ”Remote Control" menu option allows you to select
WiCo or other device as the remote control device.
The ”Bluetooth
®
” menu option allows you to activate/deac-
tivate the ”Bluetooth
®
” function. A tick (√) shows that the
option is activated.
Information:
The ”Information” option offers you an overview of the most im-
portant system settings of the stirrer.
NOTICE