16
Temperature:
In the ”Temperature” menu the user can specify that the tem-
perature sensor is displayed on the display/working screen. A tick
shows that the option is activated. The precondition for this is that
a temperature sensor is connected to the
EUROSTAR station
. If
no temperature sensor is connected, or an error is present or the
temperature rises above 350 °C, three dashes are shown as the
temperature value.
Note:
Please comply with the temperature measurement range of
the external temperature sensor specified in the "Technical data"
section. The temperature display can be in either °C or °F (see the
”Settings“ menu).
Timer:
In the ”Timer” menu the user can specify that the timer is dis-
played on the display/working screen. A tick shows that the option
is activated. This setting allows the user to specify the actual time
for the stirring procedure.
A default time can also be set for the timer. This setting allows the
user to start the stirring task for a standard time. The device stops
automatically after expiry of the set time, and the set time used for
the stirring procedure appears in the display.
Note:
The user can stop the stirring function before expiry of the
set time. In this case the countdown of the timer is interrupted.
4. Select the stirrer tool:
Select the proper the stirrer tool: R 1302 HV.
Press on "OK" to confirm the setting.
> Viskosität
Rührer
OK
R 1345 HV
> Viskosität
Rührer
OK
R 1302 HV
> Viskosität
Rührer
OK
R 1302 HV
5. Activate the viscosity display:
Viskosität
Dichte
Rührer
Display
Automat. Nullabgleich
Viskosität
Dichte
Rührer
Display
Automat. Nullabgleich
Viskosität
Dichte
Rührer
Display
Automat. Nullabgleich
Note:
after backing to the working screen, if the viscosity is not shown, please enter menu "Display" and deactivate one of
"Torque", "Temperature" or "Timer".
Then, activated the viscosity display.
6. Secure proper vessel and stirrer tool:
Switch off the device and secure stirrer tool R 1302 to the stirrer. Then, secure the 3000 ml low shape beaker (DIN 12332) to the as-
sembly.
See section "Securing".
7. Fill 1500 ml the medium to the vessel.
8. Set the speed range and test speed:
Switch on the device. Set the device to speed range I (high torque) and set the speed to 200 rpm.
9. Start measuring and read the viscosity value:
Start the measuring by pressing the knob and the device starts to run with the set speed. Read the viscosity value when the figure get
steady.
3. Set the medium density:
Set the medium density by rotating the knob.
Press on "OK" to confirm the setting.
> Viskosität
Dichte
OK
1.000
g/cm
3
> Viskosität
Dichte
OK
0.980
g/cm
3
> Viskosität
Dichte
OK
0.980
g/cm
3
Viskosität
Dichte
Rührer
Display
Automat. Nullabgleich