![IKA RV 10 auto Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/ika/rv-10-auto/rv-10-auto_operating-instructions-manual_624745023.webp)
23
Setting the lower end stop
Depending on the size of the piston,
the setting angle of the rotation drive
and the position of the heating bath
and lift, the evaporation flask can be in
contact with the bottom of the heating
bath.
Glass may break!
Limit the lower lift position with the variable end stop.
The lift can be moved to any desired position in manual mode
using the “
” and “
” keys. There is no automatic cut-off if a
collision occurs.
• Press the “
” key until the lift reaches the desired position.
Note:
The evaporation flask should be 2/3 immersed in the hea-
ting bath.
• To adjust the stop element (
1
), press the center button (
2
) on
the front of the lift.
• Move the stop (
1
) to the desired position
• Press the “
” key until the lift has reached the upper end stop.
Note:
The path is limited from 0 – 6 cm.
Check that the limit stop is functioning properly:
• Lower the lift by holding down the “
” key.
• The drive stops automatically when the preset desired lower
end position has been reached.
• Move the drive back to the upper position.
To activate the safety lift following a long period of inactivity, use
the motor to move the lift to the top or bottom position several
times prior to starting the distillation. (See the section entitled Sa-
fety Instructions - Safety Lift)!
CAUTION
CAUTION
Fill the evaporation flask
Manual filling:
You can fill the evaporation flask manually prior
to creating the vacuum. The evaporation flask should not be filled
more than half its volume.
Automatic filling:
Prior to filling the evaporation flask, a vacuum
controller is used to regulate the glass apparatus to the target
pressure.
• Now fill the evaporation flask using the back feed line.
• Due to the vacuum present, the solvent is suctioned into the
evaporation flask. This enables you to keep solvent loss due to
suctioning to a minimum.
The maximum allowed load (evapora-
tion flask and contents) is 3 kg.
Setting up the heating bath
Refer to the operating instructions
for the heating bath IKA HB!
• Move the lift to the bottom position and check the position
of the heating bath in relation to the evaporation flask. When
using larger evaporation flasks (2 or 3 litres) or depending on
the angle of the rotation drive, you can move the heating bath
50 mm to the right.
• Fill the heating bath with the tempering medium until the eva-
poration flask is surrounded by tempering medium to 2/3 of its
volume.
• Switch on the rotation drive and slowly increase the speed
Note:
Avoid creating waves.
• Switch on the heating bath using the main on/off switch.
Note:
Avoid stress on the glass due to different evaporation flask
and heating bath temperatures when lowering the evaporation
flask into the heating bath!
Note:
If non-original accessories are used that are not supplied
by
IKA
, then it is possible that the 50 mm travel range provided
by the heating bath will not be sufficient. This applies in particular
when using 3 L evaporator flasks with foam brake.
Use the
IKA
RV 10.3000 extension plate to extend the heating
bath travel range by 150 mm.
CAUTION
CAUTION
1
2
Содержание RV 10 auto
Страница 29: ...29 ZH Device setup 2 29 30 31 32 32 33 34 40 49 50 50 51 52 53 154 156 Solvent table excerpt 159...
Страница 31: ...31 IKA 100...
Страница 33: ...33 RV 10 auto RV 10 control RV 10 4002 RV 10 control VC 2 4 RV 10 auto RV 10 4002 IKA Vacstar digital RV 10...
Страница 34: ...34 30 1 30 RV 10 auto 1 140 mm 2 1 2 1...
Страница 35: ...35 1 60 60 CLICK 1...
Страница 36: ...36 2 1 3 4 1 5 6 4 6 1 2 5 6 4 3 1 2 3b 3a 3c 3a 3a 2 3b 120 rpm 20 3b 1 2 3a 3b 3c...
Страница 39: ...39 RV 10 5001 RV 10 5001 1 2 1 2 RV 10 5001 RV 10 4002 RV 10 control...
Страница 40: ...40 IKA USB USB USB PC 100 RV 10 auto...
Страница 48: ...48 2 3 1 2 1 0 6 cm 3 kg IKA HB 2 3 50 mm 2 3 IKA 50 mm 3 IKA RV 10 3000 150 mm 1 2...
Страница 50: ...50 IKA IKA IKA IKA www ika com IKA www ika com 4 IKA www ika com...
Страница 51: ...51 RV 10 4002 RV 10 4002 IN OUT VENT IN OUT VENT PCB PCB 40 C PCB...
Страница 52: ...52 PCB PC RS 232 USB LWS LWS PCB EEPROM EEPROM EEPROM IKA IKA IKA 2...
Страница 54: ...54 KO Device setup 2 54 55 56 57 57 58 59 65 74 75 75 75 77 78 154 156 Solvent table excerpt 159...
Страница 55: ...55 25 EN 61010 2 010 1 mbar UV...
Страница 56: ...56 IKA 3 1 kg 1 1200 gr 400 gr 280 gr 100 gr 1980 gr 3100 gr 1980 gr 1120 gr Power 3 1kg IKA 3 0 kg 100...
Страница 59: ...59 30 1 30 RV 10 auto 1 140 mm 2 1 2 1...
Страница 60: ...60 1 60 60 CLICK 1...
Страница 61: ...61 2 1 4 1 3 5 Velcro 6 4 Velcro 6 1 2 5 6 4 3 Vapor tube 1 2 3a 3b 3c 3a 3a 2 3b 120 rpm 20 3b 1 2 3a 3b 3c...
Страница 64: ...64 RV 10 5001 RV 10 5001 1 2 1 2 RV 10 5001 RV 10 4002 RV 10 control LOAD...
Страница 65: ...65 IKA Bluetooth RV 10 Bluetooth PC Blu etooth USB USB RV 10 USB PC 100 RV 10 RV 10 auto...
Страница 73: ...73 2 3 1 2 1 0 6 cm 3 kg IKA HB 2 3 50 mm 2 3 on off IKA 50 mm 3 L IKA RV 10 3000 150 mm 1 2...
Страница 76: ...76 RV 10 4002 RV 10 4002 PCB...
Страница 79: ...79 JA Device setup 2 79 80 81 82 82 83 84 90 99 100 100 100 102 103 154 156 Solvent table excerpt 159...
Страница 80: ...80 25 EN 61010 2 010 1 3 1 1200 gr 400 gr 280 gr 100 gr 1980 gr 3100 gr 1980 gr 1120 gr...
Страница 81: ...81 IKA 3 1 kg IKA 3 0 kg 100...
Страница 84: ...84 30 1 30 RV 10 auto 1 140 mm 2 1 1 2 1...
Страница 85: ...85 1 60 60 CLICK 1...
Страница 86: ...86 1 2 4 1 3 5 Velcro 6 4 Velcro 6 1 2 5 6 4 3 I 1 2 3b 3a 3c 3a 3a 2 3b 120 rpm 20 3b 1 2 3a 3b 3c...
Страница 89: ...89 RV 10 5001 RV 10 5001 1 2 1 2 RV 10 5001 RV 10 4002 RV 10 control LOAD...
Страница 90: ...90 IKA Bluetooth RV 10 Bluetooth PC Bluetooth USB RV 10 USB USB PC 100 RV 10 RV 10 auto...
Страница 96: ...96 5 Program 1 10 10 rpm Note Ccw cw Ccw cw Select Edit Delete Save as Edit Delete 5 Edit Insert Delete...
Страница 98: ...98 2 3 1 2 1 0 6 cm 3 kg IKA HB 50 mm 2 3 IKA 50 3 IKA RV 10 3000 150 mm 1 2...
Страница 101: ...101 RV 10 4002 RV 10 4002 IN OUT VENT I IN OUT VENT PCB...
Страница 102: ...102 PCB 40 C PCB PCB PC RS 232 USB PC LWS RV 10 RV 10 LWS PC PCB EEPROM EEPROM EEPROM IKA 24 IKA...
Страница 104: ...104 TH Device setup 2 104 105 106 107 107 108 109 115 124 125 125 126 127 128 154 156 Solvent table excerpt 159...
Страница 105: ...105 flash point 25 fire point EN 61010 2 010 fume hood gas wash bottle 1 stress tempering medium...
Страница 109: ...109 30 1 30 Drive RV 10 auto thumb screw 1 140 2 1 2 1...
Страница 110: ...110 1 60 60 CLICK 1...
Страница 114: ...114 RV 10 5001 RV 10 5001 1 2 1 2 RV 10 5001 RV 10 4002 RV 10 con trol...
Страница 115: ...115 IKA Bluetooth RV 10 Blue tooth Bluetooth USB RV 10 USB USB 100 RV 10 RV10 auto...
Страница 121: ...121 5 Program Program 1 10 1 10 10 Ccw cw Ccw cw I Select Edit Delete Save as Edit Delete 10 Back Edit Insert Delete...
Страница 123: ...123 2 3 1 2 1 0 6 1 2 3 IKA HB 2 3 50 2 3 IKA 50 3 extension plate RV 10 3000 IKA 150...
Страница 126: ...126 RV 10 4002 RV 10 4002 IN OUT VENT Venting IN OUT VENT Analog pump PCB...
Страница 127: ...127 PCB 40 PCB i PCB i RS 232 USB LWS RV 10 control RV 10 control LWS PCB EEPROM EEPROM EEPROM IKA IKA 24...