Troubleshooting

Problem

Cause

Correction

Pump stopped and cannot be 
switched on again.

Thermal cutout activated due to motor 
over-temperature.

• Manual reset is necessary. Switch off device

or isolate the equipment from mains. Wait
approx. five minutes before restarting the
device.

• Ensure pump fan is working and pump is

placed in an area with an ambient temp. of
less than or equal to 40°C

Ultimate vacuum cannot be 
reached.

Fittings or hose connection have leakage.

• Ensure that fittings are tightened properly, use

PTFE thread seal tape at thread if necessary.
Check hose connection for any cracks. Follow
instructions for “

disassembling top cover

under “

Maintenance and Cleaning

” section

to check inner hoses for leaks.

Valves not seated properly, incorrect 
orientation or damaged.

• Contact 

IKA

®

 service department.

Diaphragm not seated in bore properly or 
damaged.

• Contact 

IKA

®

 service department.

Incorrect washer configuration assembled 
during replacement of diaphragm.

• Contact 

IKA

®

 service department.

Liquid or solid parts inside the pump 
chamber

• Use gas ballast valve to purge system of

condensates. Remove the vacuum hose to the
gas inlet and open the gas ballast valve. Turn
the pump on and run it at idle.

• If problem persists, contact 

IKA

®

 service

department.

Knocking noise during pump 
operation.

Incorrect washer configuration assembled 
during replacement of diaphragm.

• Contact 

IKA

®

 service department.

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department
- Send the instrument for repair, including a short description of the fault.

Содержание MVP 10 basic

Страница 1: ...IKA MVP 10 basic MVP 10 basic 022016 20000004618...

Страница 2: ...mation which is important for proper use and ensuring that the operations of the instrument are performed efficiently and for using the instrument Failure to observe this information may result in ina...

Страница 3: ...the Accessories chapter Only authorized personnel are allowed to disassemble the appliance The device can only be disconnected from the power sup ply network by pulling out the mains plug or the conn...

Страница 4: ...ic pressure air to be pushed into the chamber The air in this chamber is then pushed out from the vacuum chamber through a different route gas outlet with the use of valves These periodic movements of...

Страница 5: ...a vacuum load device IKA MVP 10 basic vacuum pump standalone IKA MVP 10 basic Fig 2 Electrical connections Vacuum hose Mains power Recipient Outlet Inlet Pump control Vacuum Controller Magnetic Valve...

Страница 6: ...in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the equipment when cleaning Before performing a non recommended method of clean ing or decontamination the user mu...

Страница 7: ...ballast valve refer to gure below Open Gas ballast valve is open when the notch on the cap is aligned with the notch on the nozzle Gas ballast valve can be closed by rotating cap at least 90 When gas...

Страница 8: ...ks Valves not seated properly incorrect orientation or damaged Contact IKA service department Diaphragm not seated in bore properly or damaged Contact IKA service department Incorrect washer con gurat...

Страница 9: ...tor protection Thermal cutout Operation at a terrestrial altitude m max 2000 Materials in contact with medium Components Wetted parts Vacuum crater insert PTFE Carbon reinforced Vacuum crater ETFE Car...

Отзывы: