background image

13

• Place the pump on a stable and even surface before 

switching the power on. 

Inhalation of or contact with media 
such as poisonous liquids, gases, 
spray mist, vapours, dusts or biolog-

ical materials can be hazardous to user. Make sure con-
nections are tight and free from leaks when dealing with 
such media.

• The pump system is not designed to be explosion proof. 

Ensure that the temperature of the medium being evacu-
ated with the vacuum is below the ignition temperature. 
Attention must be paid to temperature rise of medium 
while being evacuated.

• Do not operate the appliance with hazardous substances or 

under water.

• Tasks with the device must only be performed when op-

eration is monitored.

• There may be electrostatic discharges between the me-

dium and the drive which could pose a direct danger.

• 

Read the operating instructions fully before starting up 

and follow the safety instructions.

• Keep the operating instructions in a place where they can 

be accessed by everyone.

• Ensure that only trained staff work with the instrument.
•  Follow the safety instructions, guidelines, occupational 

health and safety and accident prevention regulations.

• Wear your personal protective equipment in accordance 

with the hazard category of the medium to be processed. 
There may be risk from glass breakage.

Do not operate pump when front 
cover is open. This may lead to seri-
ous injury if hand is caught in mov-

ing parts as internal parts are moving at a very high speed.

• Always ensure that front cover is attached before starting 

up the device.

• Do not lift the pump if top motor cover is not tighten 

properly.

• Do not switch the pump on in the vertical position as it is 

not stable in this position.

Safety Instructions 

•  Always operate the pump in the horizontal position.

• Safe operation is only guaranteed with the accessories de-

scribed in the ”

Accessories

” chapter.

• Only authorized personnel are allowed to disassemble the 

appliance.

• The device can only be disconnected from the power sup-

ply network by pulling out the mains plug or the connec-
tor plug.

• The socket for the mains cord must be easily accessible.
• Position the positive pressure outlet of the vacuum pump 

under a fume hood.

• Refer to the operating instructions for the equipment this 

vacuum pump is connected to, e.g. rotary evaporator.

• The motor is shut down by a thermal cutout safety switch 

in the winding.
- Manual reset is necessary. Switch off device or isolate the 

equipment from mains. Wait approx. five minutes before 
restarting the device.

- Use device in conditions specified in “

Technical Data

” 

section.

• Do not use pump to generate pressure.
• Maximum pressure at gas inlet or outlet should not ex-

ceed 1 bar.

•  Use solenoid/check valves if there are several load devic-

es.

• Use only flexible hoses.

For protection of the equipment

• The voltage stated on the type plate must correspond to 

the mains voltage.

• Socket must be earthed (protective ground contact).
• Removable parts must be refitted to the appliance to pre-

vent the infiltration of foreign objects, liquids etc.

• Protect the appliance and accessories from bumps and im-

pacts.

• The instrument may only be opened by experts.
• Do not cover the ventilation slots of the instrument in 

order to ensure adequate cooling of the instrument.

• Use only genuine spare parts for maintenance to ensure 

good working condition of the instrument.

• Pay attention to water condensation in and outside of the 

instrument. Allow instrument to warm up if brought in from 
a cold environment.

• Do not use the pump to evacuate any solid substance as 

this will damage the diaphragm and the internal parts of the 
pump.

• Do not use the pump to evacuate liquids as it will reduce 

the lifespan of the diaphragm and cause leakage.

• Ensure that any substances to be used with the instru-

ment are compatible to the materials of the “

Wetted 

parts

” of the device; see section “

Technical data

”.

DANGER

DANGER

Содержание MVP 10 basic

Страница 1: ...IKA MVP 10 basic MVP 10 basic 022016 20000004618...

Страница 2: ...mation which is important for proper use and ensuring that the operations of the instrument are performed efficiently and for using the instrument Failure to observe this information may result in ina...

Страница 3: ...the Accessories chapter Only authorized personnel are allowed to disassemble the appliance The device can only be disconnected from the power sup ply network by pulling out the mains plug or the conn...

Страница 4: ...ic pressure air to be pushed into the chamber The air in this chamber is then pushed out from the vacuum chamber through a different route gas outlet with the use of valves These periodic movements of...

Страница 5: ...a vacuum load device IKA MVP 10 basic vacuum pump standalone IKA MVP 10 basic Fig 2 Electrical connections Vacuum hose Mains power Recipient Outlet Inlet Pump control Vacuum Controller Magnetic Valve...

Страница 6: ...in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the equipment when cleaning Before performing a non recommended method of clean ing or decontamination the user mu...

Страница 7: ...ballast valve refer to gure below Open Gas ballast valve is open when the notch on the cap is aligned with the notch on the nozzle Gas ballast valve can be closed by rotating cap at least 90 When gas...

Страница 8: ...ks Valves not seated properly incorrect orientation or damaged Contact IKA service department Diaphragm not seated in bore properly or damaged Contact IKA service department Incorrect washer con gurat...

Страница 9: ...tor protection Thermal cutout Operation at a terrestrial altitude m max 2000 Materials in contact with medium Components Wetted parts Vacuum crater insert PTFE Carbon reinforced Vacuum crater ETFE Car...