Use

Together with the accessories recommended by 

IKA

®

the device is suitable for:

- Evacuation of gas from a lab device, e.g. a rotary evapo-

rator.

Mode of operation: Tabletop device.

• Range of use (indoor use only)

- Laboratories   

- Schools

- Pharmacies   

- Universities

This equipment is suitable for use in all areas except: 
- Residential areas 
-  Areas that are connected directly to a low-voltage supply 

network that also supplies residential areas. 

The safety of the user cannot be guaranteed: 
- If the instrument is operated with accessories that are not sup-

plied or recommended by the manufacturer 

- If the instrument is operated improperly contrary to the 

manufacturer’s specifications 

- If the instrument or the printed circuit board are modified 

by third parties.

• Unpacking

- Please unpack the device carefully
- In the case of any damage a detailed report must be 

sent immediately (post, rail or forwarder)

Unpacking

• Delivery scope

IKA

® 

MVP 10 basic

- Operating instructions.

Diaphragm vacuum pumps (also called membrane pumps) 
are oscillating positive displacement pumps. The dia-
phragms generate vacuum by expanding the volume in the 
vacuum chamber which in turn reduces the pressure in the 
vacuum chamber thus causing atmospheric pressure air to 
be pushed into the chamber. The air in this chamber is then 
pushed out from the vacuum chamber through a different 
route (gas outlet) with the use of valves. These periodic 
movements of the diaphragm generates vacuum at the in-
let of the pump.
Attention have to be paid when working with gas mixtures 
containing condensable gases, e.g. water vapour or sol-
vents. Use of

 IKA

®

 MVP 10 basic

 with these gases will 

result in their condensing somewhere in the air flow of the 
pump. This causes pressure build up in the vacuum chamber 
which in turn reduces the service life of the diaphragms and 
valves. 

Useful information

To prevent condensation or to discharge condensates pro-
duced by the pump, the gas ballast valve should be opened. 
By opening the gas ballast valve, a small amount of air is 
allowed into the vacuum chamber, reducing the concentra-
tion of the condensable gas. Purging of condensation can 
also be done by allowing full gas flow at atmospheric pres-
sure through the pump for several minutes.

To protect the internal valves and diaphragms, fit an up-
stream condensation trap (such as a Woulff bottle) to the 

IKA

®

 rotary evaporator

 RV 8

 and 

RV10

).

The release of solvent vapors into the atmosphere can be 
avoided by fitting a downstream discharge condensation 
trap.
The exhaust air hose should generally deliver to a laboratory 
extraction system.
Fitting an exhaust air hose will also reduce the noise level.

Содержание MVP 10 basic

Страница 1: ...IKA MVP 10 basic MVP 10 basic 022016 20000004618...

Страница 2: ...mation which is important for proper use and ensuring that the operations of the instrument are performed efficiently and for using the instrument Failure to observe this information may result in ina...

Страница 3: ...the Accessories chapter Only authorized personnel are allowed to disassemble the appliance The device can only be disconnected from the power sup ply network by pulling out the mains plug or the conn...

Страница 4: ...ic pressure air to be pushed into the chamber The air in this chamber is then pushed out from the vacuum chamber through a different route gas outlet with the use of valves These periodic movements of...

Страница 5: ...a vacuum load device IKA MVP 10 basic vacuum pump standalone IKA MVP 10 basic Fig 2 Electrical connections Vacuum hose Mains power Recipient Outlet Inlet Pump control Vacuum Controller Magnetic Valve...

Страница 6: ...in the cleansing agent for the purpose of cleaning Do not allow moisture to get into the equipment when cleaning Before performing a non recommended method of clean ing or decontamination the user mu...

Страница 7: ...ballast valve refer to gure below Open Gas ballast valve is open when the notch on the cap is aligned with the notch on the nozzle Gas ballast valve can be closed by rotating cap at least 90 When gas...

Страница 8: ...ks Valves not seated properly incorrect orientation or damaged Contact IKA service department Diaphragm not seated in bore properly or damaged Contact IKA service department Incorrect washer con gurat...

Страница 9: ...tor protection Thermal cutout Operation at a terrestrial altitude m max 2000 Materials in contact with medium Components Wetted parts Vacuum crater insert PTFE Carbon reinforced Vacuum crater ETFE Car...