• Each individual instruction including parameters and data as
well as each reply are terminated with CR LF (Code: hex
0x0D and 0x0A) and have a maximum lenght of 80 charac-
ters.
• The decimal separator in a floating point number is the point.
(Code: hex 0x2E)
The above statements largely correspond with the recommen-
dations of the NAMUR-Assocation (NAMUR-recommendati-
ons for the design of electric plug connections for the analog
and digital signal transmission to labortory - MSR individual
units. Rev. 1.1).
Overview of the NAMUR-instructions
Abbreviations:
X,y =
numbering parameter (integer number)
m =
variable value, integer
n
=
value of variable, floating point number
X
= 1
Pt1000 medium temperature (external temperature
sensor)
X
= 2
temperature (incubations room)
X
= 3
safety temperature
X
= 4
speed
X
= 6
safety speed
X
= 50 Pt1000.60 medium temperature sensor offset in K
(-5.0 <= n <=+5.0)
X
= 52 incubations room temperature sensor offset in K
(-5.0 <= n <=+5.0)
NAMUR instuctions Function
IN_NAME
Input description name
IN_PV_X
Reading the real value
X=1;2;3;4;
IN_SOFTWARE
Input software ID number
date, version
IN_SP_X
Reading the set rated value
X=1;2;3;4;6;12;
42;50;52;53;
IN_TYPE
Input laboratory unit ID
OUT_NAME name Output description name. (Max. 10 characters,
default: KS4000 ic)
OUT_SP_12@n
Setting the WD safety temperature with the
echo of the set value
OUT_SP_42@n
Setting the WD safety speed with the
echo of the set value
OUT_SP_X n
Setting the rated value to n
X=1;2;4;50;52
OUT_WD1@m
Watchdog mode 1: When a WD1 event occurs,
the heating and shaking functions are shutdown
and message PC 1 is displayed.
Set the watchdog time to m (20...1500) seconds,
with echo of the watchdog time. This instruction
starts the watchdog function and must be sent
within the set watchdog time.
OUT_WD2@m
Watchdog mode 2: When a WD2 event occurs,
the speed setpoint will be set to the WD safety
setpoint speed and the temperature setpoint
will be set to the WD safety setpoint temperatu-
re. The PC 2 warning is displayed.
The WD2 event can be reset with OUT_WD2@0-
resetting also blocks the watchdog function. Set
the watchdog time to m (20...1500) secondes,
with echo of the watchdog time. This command
starts the watchdog function and must be sent
within the set watchdog time.
RESET
Switching off the instrument function.
All manuals and user guides at all-guides.com
Содержание KS 400 ic
Страница 45: ...45 KS 4000 i 1208 All manuals and user guides at all guides com...
Страница 62: ...62 KS 4000 i 1208 EL All manuals and user guides at all guides com...
Страница 63: ...63 KS 4000 i 1208 20mm 25 C IKA 100mm All manuals and user guides at all guides com...
Страница 80: ...80 KS 4000 i 1208 BG All manuals and user guides at all guides com...
Страница 81: ...81 KS 4000 i 1208 20 mm 25 IKA 100 mm All manuals and user guides at all guides com a l l g u i d e s c o m...