Technical data

Speed range 

I

 (high torque)

Speed range 

II

 (high speed)

rpm

0 / 6–400
0 / 30–2000

Speed setting

Stepless 

Speed display

7-Segment LEDs

Speed setting accuracy

rpm

± 1 

Speed measurement deviation

Speed < 300 rpm: ± 3 rpm
Speed > 300 rpm: ± 1 %

Max. torque at stirrer shaft

Ncm

Range 

I

: 400 Ncm 

Range 

II

: 80 Ncm

Max. stirring quantity (water)

ltr

150 

Max. viscosity

mPas

100000

Permitted on time

%

100 

Operating voltage

VAC

230 ± 10 % 
115 ± 10 %
100 ± 10 %

Frequency

Hz

50 / 60

Max. input power

W

226

Max. output power at stirring shaft

W

167

Protection type acc. to EN 60529

IP 40 

Protection class

I

Excess voltage category

II 

Contamination level

Protection at overload

Yes / motor current limitation 

Fuse (on mains plate)

A

T 6.3 A (

IKA

®

 Ident. No. 2585000) 

Ambient temperature

°C

+ 5 to + 40 

Ambient humidity (rel.)

%

80

Drive

Brushless motor

Clamping chuck – clamping range

mm

0.5–10

Hollow shaft internal diameter

mm

10.3

Extension arm (Ø x L)

mm

16 x 160 

Housing

Coated aluminium casting and thermoplastic plastic

Dimensions (W x D x H) (without extension arm)

mm

114 x 245 x 300

Weight (with extension arm and clamping chuck)

kg

7.8

Operation at a terrestrial altitude

m

max. 2000 

Subject to technical changes!

Содержание EUROSTAR 400 digital

Страница 1: ...20000007307 IKA EUROSTAR 400 digital Operating instructions EN EUROSTAR 400 digital_032016...

Страница 2: ...2 1 2 3 4 Fig 2 Fig 3 H I J G G Fig 4 K L M Fig 5 T U Q S U Fig 1 A B C D E F Fig 7 Fig 6 X I G Z H G W V...

Страница 3: ...bility that this product corresponds to the regulations 2014 35 EU 2006 42 EC 2014 30 EU and 2011 65 EU and conforms with the standards or other normative documents EN 61010 1 EN 61010 2 051 UL 61010...

Страница 4: ...substances or under water Process pathogenic materials only in closed vessels under a suitable fume hood Please contact IKA application support if you have any question The instrument is not suitable...

Страница 5: ...nit for safe operation in order to prevent for example the ingress of fluids for eign matter etc Wear your personal protective equipment in accordance with the hazard category of the me dium to be pro...

Страница 6: ...er shaft will go down quickly to stop Display symbol Er 4 see section Error codes Drive The speed can be adjusted continuously over two speed range speed range I speed range II with the knob B see Fig...

Страница 7: ...sary for safe use to occasionally check that the extension arm is attached securely Tighten the hexangular bolt as if required Securing the overhead stirrer to the stand Diagram see Fig 3 First secure...

Страница 8: ...aging is not sufficient Please also use suitable transport packaging Switching on the instrument Check whether the voltage given on the type plate corresponds to the available mains voltage The socket...

Страница 9: ...actory enclosing the delivery invoice and giv ing reasons for the claim You will be liable for freight costs R 2722 H stand R 2723 Telescopic stand R 2850 Floor stand R 271 Cross sleeve The warranty d...

Страница 10: ...put power at stirring shaft W 167 Protection type acc to EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload Yes motor current limitation Fuse on...

Страница 11: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: