• It is important to note that the surfaces of the motor 

(cooling fins) and certain parts of the bearing may get 
very hot during operation.

• Never cover the ventilation slots or cooling fins on the 

motor or on the instrument.

• Avoid knocking and impacting on the lower end of the 

shaft and the chuck gear teeth. Even minor, invisible dam-
age can lead to imbalance and uneven shaft action. 

• Ensure that the stand does not start to move.
• Imbalance of the output shaft, the chuck and in particu-

lar the stirring tools can lead to uncontrolled resonant 
vibrational behavior of the instrument and the whole 
assembly. Glass apparatus and stirrer containers can be 
damaged or shattered by this. It can cause injury to the 
operator, also can damage the rotating stirring tool. In 
this case exchange the stirring tool for one without im-
balance or remedy the cause of the imbalance. If there is 
still imbalance, return it to the dealer or the manufacturer 
along with a description of the fault.

• If the instrument is operated too long in overload or if the 

ambient temperature is too high, the instrument switches 
off permanently.

• The machine must only be opened by trained specialists, 

even during repair. The instrument must be unplugged 
from the power supply before opening. Live parts inside 
the instrument may still be live for some time after un-
plugging from the power supply.

Covering or parts that are ca-
pable of being removed from the 
unit without accessory equip-

ment have to be reattached to the unit for safe operation 
in order to prevent, for example, the ingress of fluids, for-
eign matter, etc..

Wear your personal protective 
equipment in accordance with 
the hazard category of the me-

dium to be processed, otherwise there is a risk of:
- splashing of liquids
- projectile parts
- body parts, hair, clothing and jewelry getting caught.

Beware of the risk of:

- flammable materials
- glass breakage as a result of mechanical shaking power.

Reduce the speed if:

-  the medium splashes out of the vessel because the 

speed is too high

-  the instrument is not running smoothly
-  the instrument begins to move around because of 

dynamic forces

-  an error occurs.

Do not touch rotating 

 

parts during operation!

• There may be electrostatic activity between the me-

dium and the output shaft which could cause a direct 
danger. 

• After an interruption in the power supply or a mechanical 

interruption during a stirring process, the unit does not 
restart automatically.

WARNING

WARNING

WARNING

NOTICE

NOTICE

Correct use

• Use: 

 

For mixing/stirring liquids with low to high viscosities by 
various stirring tools.  

Intended use: Stand device (chuck points down)

• Range of use (indoor use only):

- Laboratories   

- Schools

- Pharmacies   

- Universities

This instrument is suitable for use in all areas except: 
- Residential areas. 
- Areas that are connected directly to a low-voltage supply 

network that also supplies residential areas. 

The safety of the user cannot be guaranteed: 
- If the instrument is operated with accessories that are not 

supplied or recommended by the manufacturer. 

- If the instrument is operated improperly contrary to the 

manufacture’s specifications.

- If the instrument is operated improperly contrary to the 

manufacturer’s specifications. 

Содержание EUROSTAR 400 digital

Страница 1: ...20000007307 IKA EUROSTAR 400 digital Operating instructions EN EUROSTAR 400 digital_032016...

Страница 2: ...2 1 2 3 4 Fig 2 Fig 3 H I J G G Fig 4 K L M Fig 5 T U Q S U Fig 1 A B C D E F Fig 7 Fig 6 X I G Z H G W V...

Страница 3: ...bility that this product corresponds to the regulations 2014 35 EU 2006 42 EC 2014 30 EU and 2011 65 EU and conforms with the standards or other normative documents EN 61010 1 EN 61010 2 051 UL 61010...

Страница 4: ...substances or under water Process pathogenic materials only in closed vessels under a suitable fume hood Please contact IKA application support if you have any question The instrument is not suitable...

Страница 5: ...nit for safe operation in order to prevent for example the ingress of fluids for eign matter etc Wear your personal protective equipment in accordance with the hazard category of the me dium to be pro...

Страница 6: ...er shaft will go down quickly to stop Display symbol Er 4 see section Error codes Drive The speed can be adjusted continuously over two speed range speed range I speed range II with the knob B see Fig...

Страница 7: ...sary for safe use to occasionally check that the extension arm is attached securely Tighten the hexangular bolt as if required Securing the overhead stirrer to the stand Diagram see Fig 3 First secure...

Страница 8: ...aging is not sufficient Please also use suitable transport packaging Switching on the instrument Check whether the voltage given on the type plate corresponds to the available mains voltage The socket...

Страница 9: ...actory enclosing the delivery invoice and giv ing reasons for the claim You will be liable for freight costs R 2722 H stand R 2723 Telescopic stand R 2850 Floor stand R 271 Cross sleeve The warranty d...

Страница 10: ...put power at stirring shaft W 167 Protection type acc to EN 60529 IP 40 Protection class I Excess voltage category II Contamination level 2 Protection at overload Yes motor current limitation Fuse on...

Страница 11: ...IKA Werke GmbH Co KG Janke Kunkel Str 10 D 79219 Staufen Tel 49 7633 831 0 Fax 49 7633 831 98 sales ika de www ika com...

Отзывы: