0DOIXQFWLRQZLWKRXWPHVVDJHLQWKHGLVSOD\
(UURU
&DXVH
5HPHG\
&DORULPHWHULQXQGHILQHG
VWDWH
Power failure
If no measurement was running, the
system can be restarted by turning it
off and back on again.
If the system was in the process of
performing a measurement, the pro-
cedure is as follows:
•
Turn the system off and back on
again and start in the normal,
proper manner.
•
If residual water is left in the in-
ner vessel, discharge it with
Maintenance menu
(PSW\,9
•
Remove the decomposition ves-
sel, relieve the pressure and
prepare it again.
/RXGSHUPDQHQWKLVVLQJ
GLVFKDUJHGXULQJWKH2
ILOOLQJSURFHGXUHVHF
O
2
filling is not working
Examine:
•
the decomposition vessel to be
certain it is snugly in place (see
Section 7.4)
•
the condition of the seal on the
filling piston (see Section 11.5)
•
the seal of the decomposition
vessel C 5010/ 5012 Operating
Instructions)
,QFRPSOHWHFRPEXVWLRQ
Insufficient quantity of oxygen in
the decomposition vessel.
Sample has low flammability
Check the O
2
supply to the device
(30 bar).
Use combustion aids.
([SHULPHQWFDQQRWEH
VWDUWHG
When the decomposition vessel
is suspended into the measuring
cell, the Bomb
È
display does
not change to START.
Check the following items:
•
The decomposition vessel was
not removed after the last meas-
urement.
•
No experiment has been pre-
pared.
•
The Maintenance menu is open.
•
There is a problem with the con-
tact spring (see Section 11.6).
•
The ignition wire is defective.