2SHUDWLRQ
configuration field
[ ]
3URWRFRO
Marking this option causes a protocol to be printed out for every experiment.
This also applies to the individual temperature measurement values during the
measurement. In addition, the temperature protocol is displayed in the info-
window for test parameters, see Section 9.1, “5-Info”.
[ ]
6DPSOHUDFN
A sample rack is connected and will be used.
6DPSOH
dialog box can no longer
be called manually. Instead it is activated by placing samples in the samples
rack or removing them from it. This ensures reliable management of large num-
bers of samples.
For more information on working with the sample rack, please refer to the
C 5020 Operating Instructions.
[ ]
%RPE,G
Decomposition vessels are automatically identified by their coding (DV detec-
tion). There is no manual entry of the code number of decomposition vessels.
See Chapter 7, “Preparing and performing measurements”. This selection is de-
activated when a device without automatic identification is being run.
[ ]
6DYHH[SHULPHQW
The experiment can be started again if it is interrupted before the ignition takes
place. The experiment parameters remain intact. Even if a test is interrupted
after ignition with the error message
1RWHPSHUDWXUHLQFUHDVH
, the test can still
be restarted. To restart the experiment, the decomposition vessel must be re-
moved from the measuring cell and suspended again.
[ ]
8VUGHIQDPH
Here you can specify whether you will enter the
6DPSOHQDPH
yourself in the
6DPSOHQDPH
entry field or in the
6DPSOH
dialog box, or whether the system will
automatically assign the
6DPSOHQDPH
. If this option is not selected, the system
assigns experiment numbers in the
6DPSOHQDPH
box.
[ ]
'LVSRVDEOHFUXFLEOH
This option reduces the value for extraneous energy by 50 Joules since no cot-
ton thread is used.
[ ]
5H,QLW
Before every measurement the system tests whether cooling water flowthrough
in the calorimeter is ensured and if the cooling water temperature is correct for
the current mode. If these conditions are not satisfied, the measurement is not
enabled. This option should only be used for major fluctuations in the tempera-
ture of the cooling water or with very low frequency of measurements.
[ ]
([SORVLY
This option is only valid for the C 2000, which is equipped with the C 60 calo-
rimeter conversion kit. The standard decomposition vessels C 5010 and C 5012
must not be used in this mode. In the
([SORVLY
mode, use exclusively the IKA
High-Pressure Decomposition Vessel C 62, or a Peters bomb modified by fitting
IKA
Electrode Set C 61.
The option can only be activated if DV detection is deactivated. The activated
option allows the use of up to 20 decomposition vessels per calorimeter and
changes the procedure for preparing and performing measurements (see Sec-
tions 7.3 and 7.4).